![YOKOGAWA GS820 Скачать руководство пользователя страница 25](http://html.mh-extra.com/html/yokogawa/gs820/gs820_user-manual_911275025.webp)
Source Waveform and Sweep Mode Indicator
Displays the combination of the source waveform and sweep mode ON/OFF. The
following four combinations are available.
Indication
Source Waveform
Sweep
DC
DC
OFF
Pls
Pulse
OFF
DC Swp
DC
ON
Pls Swp
Pulse
ON
Wiring System Indicator
Displays the selected wiring system.
2W/2Wire: Local sense
4W/4Wire: Remote sense
Equation Computation Indicator
Displays “MATH” when equation computation is ON.
NULL Computation Indicator
Displays “Null” when NULL computation is ON.
Comparison Result Display
Displays the judgment result of the comparison.
High: Greater than the upper limit
In: Within the range
Low: Less than the lower limit
High Limiter and Low Limiter Values
Displays the high limiter and low limiter values.
High Limiter Indicator
Highlighted when the high limiter is activated.
Low Limiter Indicator
Highlighted when the low limiter is activated.
Source Level
Displays the current source level. VS/IS at the front indicates the source function (voltage
source or current source).
Measured Value
Displays the current measured value. VM/IM at the front indicates the measurement
function (voltage measurement or current measurement). The display indicates “-----”
when measurement is not performed and “+OVER” or “−OVER” when the measurement
is over-the-range.
Auto Range ON
Displayed when auto range is selected.
Source Range or Measurement Range
Displays the current range.
1-4
IM 765601-01E
1.3 Display Mode and Displayed Contents