Source and Measurement Timing
Basic Timing of Source and Measurement
The GS820 has two source modes: DC source and pulse source. In either mode, source
and measurement are carried out with a trigger input as the starting point. Measurement
is performed over a preset integration time after starting the source operation. After the
measurement, the auto zero measurement function, which measures the internal zero
reference and performs offset correction in real-time, can be used.
Integration
time
Measurement
delay
Source
delay
Source trigger
Source trigger
Measure
Source
Auto zero
measurement
For details on the source operation, see section 2.3, “Source.”
For details on the measurement operation, see section 2.4, “Measurement.”
Setting the Timing Using Triggers
The source trigger and measurement trigger can be selected separately. The source
trigger includes two types of constant period timers, external signal input, auxiliary
trigger, and measurement end. The measurement trigger includes source change,
sweep end, and immediate in addition to the triggers available for the source trigger. The
auxiliary trigger can be an external signal or an event by way of a program. Source delay,
measurement delay, and integration time can be set separately for each channel.
Measurement trigger
Measurement
delay 1
Source
delay 1
Measurement
delay 2
Source delay 2
Integration
time 1
Integration
time 2
Measurement trigger
Source trigger
Source trigger
Source trigger
Source trigger
Channel 1
Source
Channel 2
Source
Measure
Measure
Source trigger:
Timers 1 and 2 (constant period), external trigger, auxiliary trigger,
and measurement end
Measurement trigger: Timers 1 and 2 (constant period), source change, sweep end,
external trigger, and auxiliary trigger
For details on triggering, see section 2.6, “Triggering.”
2.2 Source Measure Function of the GS820
2-5
IM 765601-01E
Explanation of Functions