7.4
Setting the Integration Time
Procedure
1.
Press
CONFIG
in the MEASURE section of the front panel to display the CONFIG
menu.
2.
Press the
IntgTime
soft key to display the integration time setup screen.
3.
Use the
rotary knob
or
numeric keys&< >
to set the integration time.
If you use the numeric keys, press the
Enter
soft key to confirm the setting.
Explanation
Integration Time
The stability of measured values increases as the integration time is set longer. However,
the measurement time will be longer. If the integration time is set to an integer multiple
of the power line cycle (nPLC), it has an effect of eliminating the line frequency noise. To
perform a highly accurate measurement, set an integer value. For the procedure to set
the line frequency, see section 3.6, “Setting the Line Frequency.”
Selectable Range
0.001 PLC to 25 PLC (Power Line Cycles or 1 line cycle)
<<Corresponding Command Mnemonic>>
[:CHANnel<n>]:SENSe:NPLC <real number>|MINimum|MAXimum
[:CHANnel<n>]:SENSe:ITIMe <time>|MINimum|MAXimum
7-5
IM 765601-01E
Measurement
3
2
1
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
App
Index