14.2 GP-IB Interface Functions and Specifications
GP-IB Interface Functions
Listener Capability
• All of the settings that you can set with the panel
keys can be set through the GP-IB interface except
for power ON/OFF and communication parameters.
• Receives commands from a controller that request
the output of setup data, measured data, and the
like.
• Also receives status report commands.
Talker Capability
Outputs setup information, measured data, and other
information.
Note
Talk-only, listen-only, and controller functions are not
available on this instrument.
Switching between Remote and Local
Modes
When Switching from Local to Remote Mode
Receiving a REN (Remote Enable) message from the
controller when the instrument is in the local mode
causes the instrument to switch to the remote mode.
• The remote indicator (see section 1.1) is turned ON.
• All keys other than the SHIFT (LOCAL) key are
locked.
• Settings entered in local mode are retained even
when the GS820 switches to remote mode.
When Switching from Remote to Local Mode
Pressing
SHIFT (LOCAL)
when the instrument is in the
remote mode causes the instrument to switch to the
local mode.
• The remote indicator turns OFF.
• Key operations are enabled.
• Settings entered in remote mode are retained even
when the GS820 switches to local mode.
GP-IB Interface Specifications
• Electrical and Mechanical Specifications
Conforms to IEEE St’d 488-1978
• Mechanical Specifications
See the table below.
• Used Codes
ISO (ASCII) codes.
• Mode
Addressable mode.
• Address Selection
Set the address between 0 and 30 on the GP-IB
address setup screen (see section 14.4).
• Remote Mode Release
Clear remote mode by pressing
SHIFT (LOCAL)
.
However, key operations are invalid when Local
Lockout is enabled by the controller.
Function
Subset Name Description
Source handshaking
SH1
Full source handshaking
capability
Acceptor handshaking
AH1
Full acceptor
handshaking capability
Talker
T6
Basic talker capability,
serial polling, untalk
on MLA (My Listen
Address), and no talk-
only capability
Listener
L4
Basic listener capability,
unlisten on MTA (My Talk
Address), and no listen-
only capability.
Service request
SR1
Full service request
capability
Remote local
RL1
Full remote/local
capability
Parallel polling
PP0
No parallel polling
capability
Device clear
DC1
Full device clear
capability
Device trigger
DT1
Full device trigger
capability
Controller
C0
No controller capability
Electrical characteristics E1
Open collector