16-28
IM 765601-01E
[:CHANnel<n>]:SENSe:DELay/?
Function Sets the measurement delay or queries the
current setting.
Syntax
[:CHANnel<n>]:SENSe:DELay
<time>|MINimum|MAXimum
<time>
Specify any measurement delay
value.
MINimum Sets the minimum value (= 0 s).
MAXimum Sets the maximum value (= 3600 s).
[:CHANnel<n>]:SENSe:DELay?
→ <time> The present measurement delay time
Example
:SENS:DEL 2.5E-3
:CHAN2:SENS:DEL MIN
:CHAN1:SENS:DEL 1.25ms
:CHAN1:SENS:DEL?
[:CHANnel<n>]:SENSe:AVERage[:
STATe]/?
Function Turns ON/OFF the average function or queries the
current setting.
Syntax
[:CHANnel<n>]:SENSe:AVERage[:STATe]
1|0|ON|OFF
1 or ON
Turns the average function ON.
0 or OFF
Turns the average function OFF.
[:CHANnel<n>]:SENSe:AVERage[:STATe]?
→ 1 The average function is ON.
→ 0 The average function is OFF.
Example
:SENS:AVER ON
:CHAN2:SENS:AVER:STAT 0
:CHAN1:SENS:ACER:STAT?
[:CHANnel<n>]:SENSe:AVERage:COUNt/?
Function Sets the average count or queries the current
setting.
Syntax
[:CHANnel<n>]:SENSe:AVERage:COUNt
<integer>|MINimum|MAXimum
<integer>
Specify any average count.
MINimum Sets the minimum value (= 2).
MAXimum Sets the maximum value (= 256).
[:CHANnel<n>]:SENSe:AVERage:COUNt?
→ <integer> The present average count.
Example
:SENS:AVER:COUN MAX
:CHAN2:SENS:AVER:COUN 5
:CHAN1:SENS:AVER:COUN?
[:CHANnel<n>]:SENSe:ZERO:AUTO/?
Function Turns ON/OFF the auto zero function or queries
the current setting.
Syntax
[:CHANnel<n>]:SENSe:ZERO:AUTO
1|0|ON|OFF
1 or ON
Turns the auto zero function ON.
0 or OFF
Turns the auto zero function OFF.
[:CHANnel<n>]:SENSe:ZERO:AUTO?
→ 1 The auto zero function is ON.
→ 0 The auto zero function is OFF.
Example
:SENS:ZERO:AUTO ON
:CHAN2:SENS:ZERO:AUTO 0
:CHAN1:SENS:ZERO:AUTO?
[:CHANnel<n>]:SENSe:ZERO:EXECute
Function Executes zero calibration.
Syntax
[:CHANnel<n>]:SENSe:ZERO:EXECute
Example
:SENS:ZERO:EXEC
:CHAN2:SENS:ZERO:EXEC
[:CHANnel<n>]:SENSe:REMote/?
Function Turns ON/OFF the wiring system (ON for 4W and
OFF for 2W) or queries the current setting.
Syntax
[:CHANnel<n>]:SENSe:REMote 1|0|ON|OFF
1 or ON
Sets to remote sense (4W).
0 or OFF
Sets to local sense (2W).
[:CHANnel<n>]:SENSe:REMote?
→ 1 Set to remote sense (4W).
→ 0 Set to local sense (2W).
Example
:SENS:REM ON
:CHAN2:SENS:REM 0
:CHAN1:SENS:REM?
16.2 Commands