Explanation
Repeat Count
Set a value between 1 and 1000 or infinity (∞). If the repeat count is set to infinity (∞), the
GS820 repeats the sweep operation until you turn the sweep operation or output OFF.
Start Level, Stop Level, and Step Level
765601/765602
• Setting Resolution of the Start level, Stop Level, and Step level in Voltage
Source Mode
|Start Level|, |Stop Level|, and Step level
Setting Resolution
0.000 mV
≤ X ≤ 200.000 mV
1 µV
0.20000 V
< X ≤ 2.00000 V
10 µV
2.0000 V
< X ≤ 7.0000 V
100 µV
7.0000 V
< X ≤ 18.0000 V
100 µV
• Setting Resolution of the Start level, Stop Level, and Step level in Current
Source Mode
|Start Level|, |Stop Level|, and Step level
Setting Resolution
0.000 nA
≤ X ≤ 200.000 nA
1 pA
0.20000 µA < X ≤ 2.00000 µA
10 pA
2.0000 µA
< X ≤ 20.0000 µA
100 pA
20.000 µA
< X ≤ 200.000 µA
1 nA
0.20000 mA < X ≤ 2.00000 mA
10 nA
2.0000 mA < X ≤ 20.0000 mA
100 nA
20.000 mA < X ≤ 200.000 mA
1 µA
0.20000 A
< X ≤ 3.20000 A
10 µA
765611/765612
• Setting Resolution of the Start level, Stop Level, and Step level in Voltage
Source Mode
|Start Level|, |Stop Level|, and Step level
Setting Resolution
0.000 mV
≤ X ≤ 200.000 mV
1 µV
0.20000 V
< X ≤ 2.00000 V
10 µV
2.0000 V
< X ≤ 20.0000 V
100 µV
20.0000 V
< X ≤ 50.0000 V
100 µV
• Setting Resolution of the Start level, Stop Level, and Step level in Current
Source Mode
|Start Level|, |Stop Level|, and Step level
Setting Resolution
0.000 nA
≤ X ≤ 200.000 nA
1 pA
0.20000 µA < X ≤ 2.00000 µA
10 pA
2.0000 µA
< X ≤ 20.0000 µA
100 pA
20.000 µA
< X ≤ 200.000 µA
1 nA
0.20000 mA < X ≤ 2.00000 mA
10 nA
2.0000 mA < X ≤ 20.0000 mA
100 nA
20.000 mA < X ≤ 200.000 mA
1 µA
0.20000 A
< X ≤ 1.20000 A
10 µA
Note
• If the sweep count calculated from the start level, stop level, and step level exceeds 100000
points, an error occurs, and the sweep operation can be carried out only up to 100000 points.
• If the source range is set to auto range, the range may be changed during the sweep
operation, because the GS820 constantly selects the most suitable range.
• If set to fixed range, the sweep operation is carried out in the range best suited to the source
level with the highest resolution.
6.1 Setting the Linear Sweep
6-3