![YOKOGAWA GS820 Скачать руководство пользователя страница 90](http://html.mh-extra.com/html/yokogawa/gs820/gs820_user-manual_911275090.webp)
3
2
1
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
App
Index
4.4
Setting the Timer Period
Procedure
1.
Press
SHIFT+SYNC (TIMER)
to display the Timer menu.
2.
Press the
Timer1
or
Timer2
soft key.
3.
Use the
rotary knob
or
numeric keys&< >
to set the timer period.
If you use the numeric keys, press the soft key for the desired unit to confirm the value.
Synchronizing the Timers
2.
Press the
Timer Sync
soft key.
Explanation
The timers a general-purpose timers for the two channels. They can be used for the
following trigger sources.
Sweep start (section 5.6)
Source trigger (section 5.7)
Measurement trigger (section 7.6)
Selectable Range
100 µs to 3600.000000s
Synchronizing the Timers
Timer1 and Timer2 run at independent periods. If you want to run the two timers in sync,
you can align the phase.
Note
If you are using the timer period to perform a programmable sweep, set the CH1 source trigger
to Timer1 and CH2 source trigger to Timer2. This will cause the phases of Timer1 and Timer2
to be automatically reset at the sweep start of CH1 and CH2, respectively.
<<Corresponding Command Mnemonic>>
:TRIGger:TIMer1 <time>|MINimum|MAXimum
:TRIGger:TIMer2 <time>|MINimum|MAXimum
:TRIGger:TSYNc
*TRG