5.4
Selecting the Source Waveform and Source
Level
Procedure
1.
Press
CONFIG
in the SOURCE section of the front panel to display the CONFIG
menu.
Selecting the Source Waveform
2.
Press the
Shape
soft key to select DC or Pulse.
Setting the Source Level
3.
Use the
rotary knob
or
numeric keys&< >
to set the source level.
If you use the numeric keys, press the soft key for the desired unit to confirm the
setting.
Note
The source level can be changed while setting other items or while the output is ON.
Explanation
Source Waveform
There are two source modes: DC source and pulse source.
DC source mode
Pulse source mode
Source
Time
Time
Source
Source
level
Pulse base
value
Source level
Pulse width
Source delay
↑
Source
trigger
If you select the pulse source mode, set the pulse base value (see section 5.10, “Setting
the Pulse Base”) and pulse width (see section 5.11, “Setting the Pulse Width”).
5-8