Explanation
Repeat Count
Set a value between 1 and 1000 or infinity (∞). If the repeat count is set to infinity (∞), the
GS820 repeats the sweep operation until you turn the sweep operation or output OFF.
Start Level and Stop Level
765601/765602
• Setting Resolution of the Start level and Stop Level in Voltage Source Mode
|Start Level| and |Stop Level|
Setting Resolution
0.000 mV
≤ X ≤ 200.000 mV
1 µV
0.20000 V
< X ≤ 2.00000 V
10 µV
2.0000 V
< X ≤ 7.0000 V
100 µV
7.0000 V
< X ≤ 18.0000 V
100 µV
• Setting Resolution of the Start level and Stop Level in Current Source Mode
|Start Level| and |Stop Level|
Setting Resolution
0.000 nA
≤ X ≤ 200.000 nA
1 pA
0.20000 µA < X ≤ 2.00000 µA
10 pA
2.0000 µA
< X ≤ 20.0000 µA
100 pA
20.000 µA
< X ≤ 200.000 µA
1 nA
0.20000 mA < X ≤ 2.00000 mA
10 nA
2.0000 mA < X ≤ 20.0000 mA
100 nA
20.000 mA < X ≤ 200.000 mA
1 µA
0.20000 A
< X ≤ 3.20000 A
10 µA
765611/765612
• Setting Resolution of the Start level and Stop Level in Voltage Source Mode
|Start Level| and |Stop Level|
Setting Resolution
0.000 mV
≤ X ≤ 200.000 mV
1 µV
0.20000 V
< X ≤ 2.00000 V
10 µV
2.0000 V
< X ≤ 20.0000 V
100 µV
20.0000 V
< X ≤ 50.0000 V
100 µV
• Setting Resolution of the Start level and Stop Level in Current Source Mode
|Start Level| and |Stop Level|
Setting Resolution
0.000 nA
≤ X ≤ 200.000 nA
1 pA
0.20000 µA < X ≤ 2.00000 µA
10 pA
2.0000 µA
< X ≤ 20.0000 µA
100 pA
20.000 µA
< X ≤ 200.000 µA
1 nA
0.20000 mA < X ≤ 2.00000 mA
10 nA
2.0000 mA < X ≤ 20.0000 mA
100 nA
20.000 mA < X ≤ 200.000 mA
1 µA
0.20000 A
< X ≤ 1.20000 A
10 µA
Step Count
Set a value in the range from 2 to 100000.
Note
• If the start level and stop level have opposite signs, an error occurs, and the sweep
operation cannot be carried out. In addition, if the start level or stop level is zero, an error
occurs, and the sweep operation cannot be carried out.
• If the source range is set to auto range, the range may be changed during the sweep
operation, because the GS820 constantly selects the most suitable range.
• If set to fixed range, the sweep operation is carried out in the range best suited to the source
level with the highest resolution.
6.2 Setting the Log Sweep
6-7
IM 765601-01E