<6. Function block parameters and Methods>
6-25
IM 12A01A02-72E
3rd Edition : Oct. 01, 2015-00
l
FLXA21-PH
Table 6.13 Action of each parameters in failure mode related Sensor Transducer block (Classic status)
Alarm / Status
TB .status
pH1
Temperature1
ORP1
rH1
Ref.
impedance1
Empty
Write Unlocked
—
—
—
—
—
—
BAD-
Configuration
Error
Hard Write Lock
Switch OFF
—
—
—
—
—
—
Write Locked
—
—
—
—
—
—
Hard Write Lock
Switch ON
—
—
—
—
—
—
Abnormal Boot
Process
—
—
—
—
—
—
PB in O/S Mode
(AL.40)
— BAD-
Nonspecific
BAD-
Nonspecific
BAD-
Nonspecific
BAD-
Nonspecific
BAD-
Nonspecific
AMP Module
Failure2 (AL.02)
— BAD-Device
Failure
BAD-Device
Failure
BAD-Device
Failure
BAD-Device
Failure
BAD-Device
Failure
BAD-Device
Failure
Wash response
time failure
Fault
—
—
—
—
—
BAD-
Configuration
Error
Calibration Due Fault BAD-
Nonspecific
—
BAD-
Nonspecific
BAD-
Nonspecific
—
Outputs in
HOLD
Fault
—
—
—
—
—
mA output burn
high
Fault
—
—
—
—
—
mA output burn
low
Fault
—
—
—
—
—
Error in mA table Fault
—
—
—
—
—
mA calculation
error
Fault
—
—
—
—
—
mA configuration
error
Fault BAD-
Configuration
Error
BAD-
Configuration
Error
BAD-
Configuration
Error
BAD-
Configuration
Error
BAD-
Configuration
Error
Internal com.
error
Fault BAD-Device
Failure
BAD-Device
Failure
BAD-Device
Failure
BAD-Device
Failure
BAD-Device
Failure
Chksum err.
(CPU AS)
Fault BAD-Device
Failure
BAD-Device
Failure
BAD-Device
Failure
BAD-Device
Failure
BAD-Device
Failure
EEPROM err.
(CPU AS)
Fault BAD-Device
Failure
BAD-Device
Failure
BAD-Device
Failure
BAD-Device
Failure
BAD-Device
Failure
pH too high
Fault BAD-Sensor
Failure
—
—
BAD-Sensor
Failure
—
pH too low
Fault BAD-Sensor
Failure
—
—
BAD-Sensor
Failure
—
Temperature too
high
Fault BAD-Sensor
Failure
BAD-Sensor
Failure
—
BAD-Sensor
Failure
—
Temperature too
low
Fault BAD-Sensor
Failure
BAD-Sensor
Failure
—
BAD-Sensor
Failure
—
ORP too high
Fault
—
—
BAD-Sensor
Failure
—
—
ORP too low
Fault
—
—
BAD-Sensor
Failure
—
—
rH too high
Fault
—
—
—
BAD-Sensor
Failure
—
rH too low
Fault
—
—
—
BAD-Sensor
Failure
—
Impedance 1 too
high
Fault BAD-Sensor
Failure
—
BAD-Sensor
Failure
BAD-Sensor
Failure
—
Impedance 1 too
low
Fault BAD-Sensor
Failure
—
BAD-Sensor
Failure
BAD-Sensor
Failure
—
Impedance 2 too
high
Fault BAD-Sensor
Failure
—
BAD-Sensor
Failure
BAD-Sensor
Failure
—
Impedance 2 too
low
Fault BAD-Sensor
Failure
—
BAD-Sensor
Failure
BAD-Sensor
Failure
—
Temp. comp.
matrix error
Fault BAD-
Configuration
Error
—
—
—
—
Checksum error
in SENCOM
(Manufacturing
data)
Fault BAD-Sensor
Failure
BAD-Sensor
Failure
BAD-Sensor
Failure
BAD-Sensor
Failure
BAD-Sensor
Failure