<6. Function block parameters and Methods>
6-18
IM 12A01A02-72E
3rd Edition : Oct. 01, 2015-00
Error
Description and Remedy
NAMUR
NE107
category
DIAGNOSIS
(Classic)
DIAGNOSIS
(Condensed)
DIAGNOSIS
EXTENSION
Checksum error
Contact your local sales office.
F
Hardware
failure
electronics.
Failure of
the device or
armature
Electric
Failure
Sensor type not
correct
SENCOM sensor type is not
correct.
Change SENCOM sensor type
on the display of FLXA21.
F
Configuration
not valid
Failure of
the device or
armature
Configuration
Error
Internal error in
SENCOM
Replace SENCOM sensor.
F
Hardware
failure
electronics.
Failure of
the device or
armature
Sensor
Electric
Failure
SENCOM not
connected
SENCOM sensor is not
connected.
Check connection of SENCOM
sensor.
F
Hardware
failure
electronics.
Failure of
the device or
armature
Electric
Failure
EEPROM error
Write or read error in the
EEPROM.
Contact your local sales office.
F
Hardware
failure
electronics.
Failure of
the device or
armature
Electric
Failure
Sens. mod. not work Contact your local sales office.
F
Hardware
failure
electronics.
Failure of
the device or
armature
Electric
Failure
Table 6.6 DEVICE_STATUS_4 (FLXA21-PH: System Warning)
Error
Description and Remedy
NAMUR
NE107
category
DIAGNOSIS
(Classic)
DIAGNOSIS
(Condensed)
DIAGNOSIS
EXTENSION
Wash response time
failure
Half-value recovery time too long.
Check cleaning system. Adjust
timing parameters. Replace
measuring sensor.
—
—
—
—
Log Book (almost)
full
Logbook is more than 95% full.
Erase logbooks. Turn logbook
"full" warning off.
M
Maintenance
required
Maintenance
required
Warning for
Logbook
Calibration Due
System was not maintained
within the preset period. Perform
maintenance. Increase Calibr.
Interval.
M
Maintenance
required
Maintenance
required
Warning for
Calibration
Due