![YOKOGAWA FLXA21 Скачать руководство пользователя страница 54](http://html.mh-extra.com/html/yokogawa/flxa21/flxa21_user-manual_911262054.webp)
<6. Function block parameters and Methods>
6-28
IM 12A01A02-72E
3rd Edition : Oct. 01, 2015-00
Table 6.14 Action of each parameters in failure mode related Sensor Transducer block (Condensed
status)
Alarm / Status
TB .status
pH1
Temperature1
ORP1
rH1
Ref.
impedance1
Empty
Write Unlocked
—
—
—
—
—
—
UNCERTAIN-
initial value
Hard Write Lock
Switch OFF
—
—
—
—
—
—
Write Locked
—
—
—
—
—
—
Hard Write Lock
Switch ON
—
—
—
—
—
—
Abnormal Boot
Process
—
—
—
—
—
—
PB in O/S Mode
(AL.40)
— BAD-passivated (diagnostis alerts inhibited)
AMP Module
Failure2 (AL.02)
— BAD-maintenance alarm, more diagnosis available
Wash response
time failure
Fault
—
—
—
—
—
UNCERTAIN-
initial value
Calibration Due Fault GOOD-
maintenance
demanded
—
GOOD-
maintenance
demanded
GOOD-
maintenance
demanded
—
Outputs in
HOLD
Fault
—
—
—
—
—
mA output burn
high
Fault
—
—
—
—
—
mA output burn
low
Fault
—
—
—
—
—
Error in mA table Fault
—
—
—
—
—
mA calculation
error
Fault
—
—
—
—
—
mA configuration
error
Fault BAD-maintenance alarm, more diagnosis available
Internal com.
error
Fault BAD-maintenance alarm, more diagnosis available
Chksum err.
(CPU AS)
Fault BAD-maintenance alarm, more diagnosis available
EEPROM err.
(CPU AS)
Fault BAD-maintenance alarm, more diagnosis available
pH too high
Fault BAD-
maintenance
alarm, more
diagnosis
available
—
—
BAD-
maintenance
alarm, more
diagnosis
available
—
pH too low
Fault BAD-
maintenance
alarm, more
diagnosis
available
—
—
BAD-
maintenance
alarm, more
diagnosis
available
—
Temperature too
high
Fault BAD-
maintenance
alarm, more
diagnosis
available
BAD-
maintenance
alarm, more
diagnosis
available
—
BAD-
maintenance
alarm, more
diagnosis
available
—
Temperature too
low
Fault BAD-
maintenance
alarm, more
diagnosis
available
BAD-
maintenance
alarm, more
diagnosis
available
—
BAD-
maintenance
alarm, more
diagnosis
available
—
ORP too high
Fault
—
—
BAD-
maintenance
alarm, more
diagnosis
available
—
—
ORP too low
Fault
—
—
BAD-
maintenance
alarm, more
diagnosis
available
—
—