<6. Function block parameters and Methods>
6-15
IM 12A01A02-72E
3rd Edition : Oct. 01, 2015-00
Table 6.3 DEVICE_STATUS_1
Error
Description and Remedy
NAMUR
NE107
category
DIAGNOSIS
(Classic)
DIAGNOSIS
(Condensed)
DIAGNOSIS
EXTENSION
Write Unlocked
Writing to parameters is
Unlocked.
Change the PB Write Lock(PB.
WRITE_LOCK) to Locked or turn
on the hardware write lock switch.
—
—
—
—
Hard Write Lock SW
OFF
Hardware write lock switch is
OFF.
Turn on the hardware write lock
switch.
—
—
—
—
Write Locked
Writing to parameters is locked.
Change the PB Write Lock(PB.
WRITE_LOCK) to Locked or turn
off the hardware write lock switch.
—
—
—
—
Hard Write Lock SW
OFF
Hardware write lock switch is ON.
Turn off the hardware write lock
switch.
—
—
—
—
Abnormal Boot
Process
Abnormal boot processing was
detected at the starting.
Check the cables and power.
F
—
Failure of
the device or
armature
—
PB in O/S Mode
(AL.40)
Physical Block is in O/S mode.
Change the PB Block Mode.
Target(PB.TARGET_MODE) to
Auto mode.
F
—
—
Function
Block Mode
Check
AMP Module
Failure2 (AL.02)
Amplifier EEPROM failed.
Replace electrical parts such
as the amplifier. Or replace the
device.
F
Hardware
failure
electronics.
Failure of
the device or
armature
Electric
Failure