![YOKOGAWA FLXA21 Скачать руководство пользователя страница 16](http://html.mh-extra.com/html/yokogawa/flxa21/flxa21_user-manual_911262016.webp)
<3. Getting Started>
3-3
IM 12A01A02-72E
3rd Edition : Oct. 01, 2015-00
3.5
Notes When Using the FLXA21-SC
Concentration Table (when using FieldMate)
The FLXA21-SC concentration table can be read from and written to using the CONC_TABLE_
CONCENTRATION_1 to CONC_TABLE_CONCENTRATION_21 and CONC_TABLE_
CONDUCTIVITY_1 to CONC_TABLE_CONDUCTIVITY_21 parameters of the transducer block.
However, when using FieldMate, note the following points.
The default concentration table values are “Not a Number”. On the device screen, they appear as
blank. On FieldMate, “IEEE: NOT A NUMBER” is displayed.
Values set to “Not a Number” cannot be changed from FieldMate.
If you want to write the concentration table from FieldMate, first set all the items in the
concentration table to some value from the device screen.
3.6 PRIMARY/SECONDARY/TERTIARY/
QUATERNARY_VALUE Value Assignment
Measurement values are assigned to PRIMARY_VALUE, SECONDARY_VALUE, TERTIARY_
VALUE, and QUATERNARY_VALUE from the device screen.
To set PRIMARY_VALUE, choose Commissioning > Output setup > Output, and then set the
Process parameter item on the mA (Output) screen.
For others, choose Commissioning > Advanced setup > Communication > HART, and on the
HART setup screen, set SECONDARY_VALUE with the SV item, TERTIARY_VALUE with the
TV item, and QUATERNARY_VALUE with the QV item.
CAUTION
Be sure to use the default values for the following settings.
Changing them may disrupt communication.
Commissioning > Advanced setup > Communication screen
Default value: HART
Commissioning > Advanced setup > Communication > HART setup screen
Item name:
Network address
Default value: 0
If you want to change the settings from the device, first change Transducer Block Mode (refer
also to tables in chapter 6 because not all parameters need OS to allow a change) to Out of
Service.
If you want to load the factory settings or change the language, first change Physical Block Mode
(refer also to tables in chapter 6 because not all parameters need OS to allow a change) to Out of
Service.
Note that when you do, the FLXA21 will restart.