<6. Function block parameters and Methods>
6-20
IM 12A01A02-72E
3rd Edition : Oct. 01, 2015-00
Error
Description and Remedy
NAMUR
NE107
category
DIAGNOSIS
(Classic)
DIAGNOSIS
(Condensed)
DIAGNOSIS
EXTENSION
Temp. comp. matrix
error
Temp. comp. matrix is not
properly defined.
Execute 'Check values' on the
display of FLXA21.
F
Configuration
not valid
Failure of
the device or
armature
Configuration
Error
Table 6.8 DEVICE_STATUS_6
Status
Description and Remedy
NAMUR
NE107
category
DIAGNOSIS
(Classic)
DIAGNOSIS
(Condensed)
DIAGNOSIS
EXTENSION
HART failure
Check HART settings on the
display of FLXA21. Contact your
local sales office.
F
Hardware
failure
electronics.
Failure of
the device or
armature
Electric
Failure
Maintenance
Someone maintenances on the
display of FLXA21.
Check it.
C
—
Function
check
Local
Operation
Table 6.9 DEVICE_STATUS_8
Status
Description and Remedy
NAMUR
NE107
category
DIAGNOSIS
(Classic)
DIAGNOSIS
(Condensed)
DIAGNOSIS
EXTENSION
TB in O/S Mode
Blank
or
Confirm your settings.
F
—
—
Function Block
Mode Check
AI3 Empty
F
—
—
Empty(value
invalid)
AI3 Hi Hi Alarm
—
—
—
AI3 HI HI/LO LO
Alarm
AI3 Hi Alarm
—
—
—
—
AI3 Lo Alarm
—
—
—
—
AI3 Lo_Lo Alarm
—
—
—
AI3 HI HI/LO LO
Alarm
AI3 Simulate Active
C*
—
—
AI Simulate
Active
AI3 in MAN Mode
C*
—
—
AI Simulate
Active
AI3 in O/S Mode
F
—
—
Function Block
Mode Check
AI2 Empty
F
—
—
Empty(value
invalid)
AI2 Hi Hi Alarm
—
—
—
AI2 HI HI/LO LO
Alarm
AI2 Hi Alarm
—
—
—
—
AI2 Lo Alarm
—
—
—
—
AI2 Lo_Lo Alarm
—
—
—
AI2 HI HI/LO LO
Alarm
AI2 Simulate Active
C*
—
—
AI Simulate
Active
AI2 in MAN Mode
C*
—
—
AI Simulate
Active
AI2 in O/S Mode
F
—
—
Function Block
Mode Check
AI1 Empty
F
—
—
Empty(value
invalid)
AI1 Hi Hi Alarm
—
—
—
AI1 HI HI/LO LO
Alarm
AI1 Hi Alarm
—
—
—
—
AI1 Lo Alarm
—
—
—
—
AI1 Lo_Lo Alarm
—
—
—
AI1 HI HI/LO LO
Alarm
AI1 Simulate Active
C*
—
—
AI Simulate
Active
AI1 in MAN Mode
C*
—
—
AI Simulate
Active
AI1 in O/S Mode
F
—
—
Function Block
Mode Check
*:
This bit is Off 10 seconds after On.