<6. Function block parameters and Methods>
6-9
IM 12A01A02-72E
3rd Edition : Oct. 01, 2015-00
(2) Transducer block parameters FLXA21-SC
Relative
Index
Index
Parameter
Write
Mode
Initial Value
Explanation
0
16 BLOCK_OBJECT
RO
—
Information on this block such as Block Tag, DD
Revision, Execution Time etc.
1
17 ST_REV
RO
0
The revision level of the static data associated
with the function block. The revision value will be
incremented each time a static parameter value in
the block is changed
2
18 TAG_DESC
AUTO
Blank
The user description of the intended application of
the block.
3
19 STRATEGY
AUTO
0
The strategy field can be used to identify grouping
of blocks.
This data is not checked or processed by the block
4
20 ALERT_KEY
AUTO
0
The identification number of the plant unit. This
information may be used in the host for sorting
alarms, etc.
5
21 TARGET_MODE AUTO
AUTO
Set the Target of block mode (MODE_BLK) to
Auto or O/S according to the write Mode of the
parameter to be set or changed.
The permitted bit is only available.
6
22 MODE_BLK
RO
AUTO,
AUTO | O/S,
AUTO
The mode parameter is a structured parameter
composed of the actual mode, the normal mode,
and the permitted mode
7
23 ALARM_SUM
RO
—
The current alarm status associated with the
function block
8
24 COMPONENT_
NAME
O/S
Blank
Description of the measurement value as readable
ASCII text.
9
25 PV
RO
—
Same as primary value.
10
26 PV_UNIT
O/S
S/cm
Unit of PV.
11
27 PV_UNIT_TEXT
O/S
Blank
Additional manufacturer specific engineering units.
12
28 ACTIVE_RANGE
O/S
RANGE_1
Number of the active range. Valid value is 1 only.
13
29 AUTORANGE_ON O/S
TRUE
Valid value is "On" only.
14
30 SAMPLING_RATE O/S
2500
Not used.
25
41 NUMBER_OF_
RANGE
RO
1
The number of ranges
26
42 RANGE_1
O/S
-3.4+038,
3.4+038
Not used
27
43 PRIMARY_
VALUE_TYPE
RO
Conductivity1-
TC1
Item of Primary value.
28
44 PRIMARY_VALUE RO
—
Primary value
29
45 PRIMARY_
VALUE_UNIT
RO
S/cm
Unit of Primary Value
30
46 SENSOR_TYPE_
SC
RO
2 electrode
Sensor type
31
47 CELL_CONST_
FACTORY
RO
0.10
Cell constant(factory setting).
32
48 CELL_CONST_
ADJUST
RO
0.10
Cell constant(adjusted)
33
49 MEASURING_
TYPE
RO
Conductivity Measuring type.
34
50 MEASURING_
UNIT
RO
/cm
Measuring unit
35
51 SENSOR_
CALIBRATION_
DATE
RO
0000/01/01
0:00:00
Date on which the last sensor calibration was
performed.
36
52 SENSOR_
CALIBRATION_
DUE_DATE
RO
0000/01/01
0:00:00
Date when the calibration must be done next.
37
53 SENSOR_TEMP_
COMPENSATION
O/S
Automatic
Temperature compensation method.
38
54 SENSOR_TEMP_
MANUAL_VALUE
O/S
25.0
Temperature used when temperature
compensation method is Manual.
39
55 REFERENCE_
TEMP
O/S
25.0
Temperature to which the measured value must be
compensated.
40
56 TEMP_
COMPENSATION1
O/S
NaCl
Temperature compensation method 1
41
57 TEMP_
COMPENSATION2
O/S
None
Temperature compensation method 2
42
58 TEMP_
COEFFICIENT1
O/S
2.10
Available on TEMP_COMPENSATION1 being TC.