<6. Function block parameters and Methods>
6-6
IM 12A01A02-72E
3rd Edition : Oct. 01, 2015-00
Relative
Index
Index
Parameter
Write
Mode
Initial Value
Explanation
43
59 REFERENCE_
TEMP
O/S
25.0
Temperature to which the measured pH value
must be compensated.
44
60 PROCESS_
TEMP_
COMPENSATION
O/S
None
Method of process temperature compensation.
45
61 PH_TEMP_
COEFFICIENT
O/S
0.000000
Coefficient of TC (Linear compensation function)
46
62 SECONDARY_
VALUE_TYPE
RO
temperature Item of Secondary value.
47
63 SECONDARY_
VALUE
RO
—
Secondary value.
48
64 SECONDARY_
VALUE_UNIT
RO
degC
Unit of secondary value
49
65 SENSOR_TYPE_
TEMP
RO
Pt1000
Temperature sensor
50
66 TEMP_UNIT
RO
degC
Unit of temperature
51
67 TERTIARY_
VALUE_TYPE
RO
Empty
Item of Tertiary value
52
68 TERTIARY_
VALUE
RO
—
Tertiary value
53
69 TERTIARY_
VALUE_UNIT
RO
none
Unit of Tertiary value
54
70 ORP_ZERO
O/S
0.000
Calculated sensor offset of ORP
55
71 ORP_SLOPE
RO
1000.0
Calibrated efficiency of ORP sensor.
56
72 QUATERNARY_
VALUE_TYPE
RO
Empty
Item of Quaternary value
57
73 QUATERNARY_
VALUE
RO
—
Quaternary value
58
74 QUATERNARY_
VALUE_UNIT
RO
none
Unit of Quaternary value
59
75 SENSOR_MV
RO
—
Voltage from sensor.
60
76 ORP_SENSOR_
MV
RO
—
Voltage from sensor for ORP.
61
77 IMPEDANCE1
RO
—
Electrical resistance of Input1.
62
78 IMPEDANCE2
RO
—
Electrical resistance of Input2.
63
79 DETC_
WELLNESS_
ZERO
RO
—
Sensor wellness indicator by Zero value.
64
80 DETC_
WELLNESS_
SLOPE
RO
—
Sensor wellness indicator by Slope value.
65
81 DETC_
WELLNESS_
IMPEDANCE1
RO
—
Sensor wellness indicator by Input1's impedance.
66
82 DETC_
WELLNESS_
IMPEDANCE2
RO
—
Sensor wellness indicator by Input1's impedance.
67
83 DETC_
WELLNESS_
HEAT_CYCLE
RO
—
Sensor wellness indicator by heat cycle.
68
84 DETC_
WELLNESS_
PROG_TIME
RO
—
Sensor wellness indicator by elapsed time.
69
85 MODULE_PDN
RO
Product no.
Product number of Sensor module.
70
86 MODULE_
SOFTREV
RO
"Rn.nn"
Software revision of Sensor module.
71
87 HOUSING_PDN
RO
Product no.
Product number of Housing module.
72
88 HOUSING_
SOFTREV
RO
“Rn.nn”
Software revision of Housing module.
73
89 SENSOR_TYPE_
MODEL
RO
—
Analog (1), SENCOM (2)
74
90 SENCOM_MAX_
TEMP
RO
—
Max temperature sensor has been exposed.
Available on SENCOM.
75
91 SENCOM_HIGH_
PH_TOTAL
RO
—
Total time during which pH value has been higher
than upper limit. Available on SENCOM.
76
92 SENCOM_LOW_
PH_TOTAL
RO
—
Total time during which pH value has been higher
than upper limit. Available on SENCOM.
77
93 SENCOM_
STERILIZATION
RO
—
Number of heat sterilization judged by preset
temperature and time. Available on SENCOM.