<6. Function block parameters and Methods>
6-3
IM 12A01A02-72E
3rd Edition : Oct. 01, 2015-00
Relative
Index
Index
Parameter
Write
Mode
Initial Value
Explanation
22
38 DEVICE_INSTAL_
DATE
AUTO
Blank
Date of installation of the device.
24
40 IDENT_
NUMBER_
SELECTOR
AUTO
Adaptation
Mode
The parameter to select IDENT Number.
Profile specific (0)
Manufacturer specific (IDENT Number of
DEVICE_ID) (1)
Adaptation Mode (127)
26
42 FEATURE
RO
—
Indicates optional feature implemented in the
device and the status of these features which
indicates if the feature is supported or not
supported.
27
43 COND_STATUS_
DIAG
AUTO
Condensed
Status and
Diagnosis
information is
provided
Indicates the mode of a device that can be
configured for stats and diagnostic behaviour.
Status and Diagnosis (0)
Condensed Status and Diagnosis information is
provided (1)
36
52 DEVICE_
CONFIGURATION
RO
Blank
Not used for FLXA21
37
53 INIT_STATE
AUTO
2
Not used for FLXA21
38
54 DEVICE_STATE
AUTO
2
Not used for FLXA21
39
55 GLOBAL_STATUS RO
—
Not used for FLXA21
50
66 SOFT_DESC
RO
—
Yokogawa internal use only
51
67 DEVICE_
STATUS_1
RO
—
Device status
52
68 DEVICE_
STATUS_2
RO
—
Device status
53
69 DEVICE_
STATUS_3
RO
—
Device status
54
70 DEVICE_
STATUS_4
RO
—
Device status
55
71 DEVICE_
STATUS_5
RO
—
Device status
56
72 DEVICE_
STATUS_6
RO
—
Device status
57
73 DEVICE_
STATUS_7
RO
—
Device status
58
74 DEVICE_
STATUS_8
RO
—
Device status
59
75 IDENT_NUMBER
RO
—
IDENT Number of the device
70
86 PRIVATE_1
RO
—
Not used for the transmitter
71
87 PRIVATE_2
RO
—
Not used for the transmitter
72
88 PRIVATE_3
RO
—
Not used for the transmitter
73
89 PRIVATE_4
RO
—
Not used for the transmitter
74
90 PRIVATE_5
RO
—
Not used for the transmitter
75
91 PRIVATE_6
RO
—
Not used for the transmitter
76
92 PRIVATE_7
RO
—
Not used for the transmitter
77
93 PRIVATE_8
RO
—
Not used for the transmitter
78
94 PRIVATE_9
RO
—
Not used for the transmitter
79
95 PRIVATE_10
RO
—
Not used for the transmitter
80
96 PRIVATE_11
RO
—
Not used for the transmitter