<6. Function block parameters and Methods>
6-27
IM 12A01A02-72E
3rd Edition : Oct. 01, 2015-00
Alarm / Status
TB .status
pH1
Temperature1
ORP1
rH1
Ref.
impedance1
Empty
Impedance 2 too
high
Warning
UNCERTAIN-
sensor
conversion
not accurate
—
UNCERTAIN-
sensor
conversion
not accurate
UNCERTAIN-
sensor
conversion
not accurate
—
BAD-
Configuration
Error
Impedance 2 too
low
Warning
UNCERTAIN-
sensor
conversion
not accurate
—
UNCERTAIN-
sensor
conversion
not accurate
UNCERTAIN-
sensor
conversion
not accurate
—
SENCOM
Logbook full
Warning
—
—
—
—
—
SENCOM
sensor changed
Warning
—
—
—
—
—
Temp. comp.
matrix error
Warning
UNCERTAIN-
Configuration
Error
—
—
—
—
SENCOM
initializing
Warning
BAD-
Nonspecific
—
BAD-
Nonspecific
BAD-
Nonspecific
—
HART failure
— BAD-Device
Failure
BAD-Device
Failure
BAD-Device
Failure
BAD-Device
Failure
BAD-Device
Failure
BAD-Device
Failure
Maintenance
—
—
—
—
—
—
BAD-
Configuration
Error
TB OOS
— BAD-out of
service
BAD-out of
service
BAD-out of
service
BAD-out of
service
BAD-out of
service
BAD-out of
service
AI3 Empty
—
—
—
—
—
—
BAD-
Configuration
Error
AI3 Hi Hi Alarm
—
—
—
—
—
—
AI3 Hi Alarm
—
—
—
—
—
—
AI3 Lo Alarm
—
—
—
—
—
—
AI3 Lo_Lo Alarm
—
—
—
—
—
—
AI3 SIM
—
—
—
—
—
—
AI3 MAN
—
—
—
—
—
—
AI3 OOS
—
—
—
—
—
—
AI2 Empty
—
—
—
—
—
—
AI2 Hi Hi Alarm
—
—
—
—
—
—
AI2 Hi Alarm
—
—
—
—
—
—
AI2 Lo Alarm
—
—
—
—
—
—
AI2 Lo_Lo Alarm
—
—
—
—
—
—
AI2 SIM
—
—
—
—
—
—
AI2 MAN
—
—
—
—
—
—
AI2 OOS
—
—
—
—
—
—
AI1 Empty
—
—
—
—
—
—
AI1 Hi Hi Alarm
—
—
—
—
—
—
AI1 Hi Alarm
—
—
—
—
—
—
AI1 Lo Alarm
—
—
—
—
—
—
AI1 Lo_Lo Alarm
—
—
—
—
—
—
AI1 SIM
—
—
—
—
—
—
AI1 MAN
—
—
—
—
—
—
AI1 OOS
—
—
—
—
—
—