![YOKOGAWA FLXA21 Скачать руководство пользователя страница 58](http://html.mh-extra.com/html/yokogawa/flxa21/flxa21_user-manual_911262058.webp)
<6. Function block parameters and Methods>
6-32
IM 12A01A02-72E
3rd Edition : Oct. 01, 2015-00
l
FLXA21-SC
Table 6.15 Action of each parameters in failure mode related Sensor Transducer block (Classic status)
Alarm / Status
TB .status
Conduct1-
TC1
Resist1-TC1
Conduct1-
TC2
Resist1-TC2
Temperature1
Concent1-
TC1
Concent1-
TC2
USP1
Empty
Write Unlocked
—
—
—
—
—
—
—
BAD-
Configuration
Error
Hard Write Lock
Switch OFF
—
—
—
—
—
—
—
Write Locked
—
—
—
—
—
—
—
Hard Write Lock
Switch ON
—
—
—
—
—
—
—
Abnormal Boot
Process
—
—
—
—
—
—
—
PB in O/S Mode
(AL.40)
— BAD-Nonspecific
AMP Module
Failure2 (AL.02)
— BAD-Device Failure
Calibration Due Fault BAD-
Nonspecific
BAD-
Nonspecific
—
BAD-
Nonspecific
BAD-
Nonspecific
BAD-
Nonspecific
BAD-
Configuration
Error
Outputs in
HOLD
Fault
—
—
—
—
—
—
mA output burn
high
Fault
—
—
—
—
—
—
mA output burn
low
Fault
—
—
—
—
—
—
Error in mA table Fault
—
—
—
—
—
—
mA calculation
error
Fault
—
—
—
—
—
—
mA configuration
error
Fault BAD-Configuration Error
Internal com.
error
Fault BAD-Device Failure
Chksum err.
(CPU AS)
Fault BAD-Device Failure
EEPROM err.
(CPU AS)
Fault BAD-Device Failure
Conductivity too
high
Fault BAD-
Sensor
Failure
BAD-
Sensor
Failure
—
BAD-
Sensor
Failure
BAD-
Sensor
Failure
BAD-
Sensor
Failure
Conductivity too
low
Fault BAD-
Sensor
Failure
BAD-
Sensor
Failure
—
BAD-
Sensor
Failure
BAD-
Sensor
Failure
BAD-
Sensor
Failure
Temperature too
high
Fault BAD-Sensor Failure
Temperature too
low
Fault BAD-Sensor Failure
Measurement
unstable
Fault BAD-Sensor Failure
Polarization
detected
Fault BAD-
Nonspecific
BAD-
Nonspecific
—
BAD-
Nonspecific
BAD-
Nonspecific
BAD-
Sensor
Failure
USP limit
exceeded
Fault BAD-
Nonspecific
BAD-
Nonspecific
—
BAD-
Nonspecific
BAD-
Nonspecific
BAD-
Nonspecific
USP margin
exceeded
Fault BAD-
Nonspecific
BAD-
Nonspecific
—
BAD-
Nonspecific
BAD-
Nonspecific
BAD-
Nonspecific
Conc. table error Fault
—
—
—
BAD-
Configuration
Error
—
—
2nd comp.
matrix error
Fault BAD-
Configuration
Error
BAD-
Configuration
Error
—
BAD-
Configuration
Error
BAD-
Configuration
Error
—
1st. comp.
matrix error
Fault BAD-
Configuration
Error
BAD-
Configuration
Error
—
BAD-
Configuration
Error
BAD-
Configuration
Error
—
Checksum error Fault BAD-Device Failure
EEPROM error Fault BAD-Device Failure
Sens. mod. not
work
Fault BAD-Device Failure