![YOKOGAWA FLXA21 Скачать руководство пользователя страница 38](http://html.mh-extra.com/html/yokogawa/flxa21/flxa21_user-manual_911262038.webp)
<6. Function block parameters and Methods>
6-12
IM 12A01A02-72E
3rd Edition : Oct. 01, 2015-00
Relative
Index
Index
Parameter
Write
Mode
Initial Value
Explanation
111
127 ERR_CONFIG_
COND_OR_
CONC_TOO_
LOW
O/S
Warning
Category of error status(Fault/Warning/Off)
112
128 ERR_CONFIG_
TEMP_TOO_
HIGH
O/S
Warning
Category of error status(Fault/Warning/Off)
113
129 ERR_CONFIG_
TEMP_TOO_LOW
O/S
Warning
Category of error status(Fault/Warning/Off)
114
130 ERR_CONFIG_
POLARIZATION_
DETECT
O/S
Warning
Category of error status(Fault/Warning/Off)
115
131 ERR_CONFIG_
CALIB_TIME_
EXCEEDED
O/S
Off
Category of error status(Fault/Warning/Off)
116
132 ERR_CONFIG_
USP_LMT_
EXCEED
O/S
Off
Category of error status(Fault/Warning/Off)
118
134 ERR_
CONFIG_1ST_
COMP_MATRIX
O/S
Fault
Category of error status(Fault/Warning/Off)
119
135 ERR_
CONFIG_2ND_
COMP_MATRIX
O/S
Fault
Category of error status(Fault/Warning/Off)
120
136 ERR_CONFIG_
CONC_TABLE
O/S
Fault
Category of error status(Fault/Warning/Off)
121
137 MEASUREMENT_
HIGH_LMT
O/S
0.250
High limit for checking "Conductivity too high" or
low limit for checking "Resistivity too low".
122
138 MEASUREMENT_
LOW_LMT
O/S
0.000001
Low limit for checking "Conductivity too low" or
high limit for checking "Resistivity too high".
123
139 USP_SAFETY_
MARGIN
O/S
0.000
Percentage of the limit value of USP<645> serves
as a safety margin.
124
140 DIAG_SETTING_
PROG_TIME
O/S
Disable
Enable or Disable of sensor wellness by elapsed
time.
125
141 DIAG_SETTING_
PROG_TIME_
BAD_LMT
O/S
2000
Limit for sensor wellness by elapsed time.
126
142 DIAG_SETTING_
HEAT_CYCLE
O/S
Disable
Enable or Disable of sensor wellness by heat
cycle.
127
143 DIAG_SETTING_
HEAT_CYCLE_
BAD_LMT
O/S
500
Limit for sensor wellness by heat cycle.
128
144 HEAT_CYCLE_
TEMP
O/S
50
Limit of temperature for sensor wellness by heat
cycle.
129
145 HEAT_CYCLE_
TIME
O/S
10.0
Limit of time for sensor wellness by heat cycle.
130
146 TRANSMITTER_
TIME
O/S
—
Time of Housing module.
144
160 TEST_1
—
—
Used by a vender’s serviceperson
DTM doesn’t support them.
145
161 TEST_2
—
—
146
162 TEST_3
—
—
147
163 TEST_4
—
—
148
164 TEST_5
—
—
149
165 TEST_6
—
—
150
166 TEST_7
—
—
151
167 TEST_8
—
—
152
168 TEST_9
—
—