![YOKOGAWA FLXA21 Скачать руководство пользователя страница 60](http://html.mh-extra.com/html/yokogawa/flxa21/flxa21_user-manual_911262060.webp)
<6. Function block parameters and Methods>
6-34
IM 12A01A02-72E
3rd Edition : Oct. 01, 2015-00
Table 6.16 Action of each parameters in failure mode related Sensor Transducer block (Condensed
status)
Alarm / Status
TB .status
Conduct1-
TC1
Resist1-TC1
Conduct1-
TC2
Resist1-TC2
Temperature1
Concent1-
TC1
Concent1-
TC2
USP1
Empty
Write Unlocked
—
—
—
—
—
—
—
BAD-
passivated
(diagnostis
alerts
inhibited
Hard Write Lock
Switch OFF
—
—
—
—
—
—
—
Write Locked
—
—
—
—
—
—
—
Hard Write Lock
Switch ON
—
—
—
—
—
—
—
Abnormal Boot
Process
—
—
—
—
—
—
—
PB in O/S Mode
(AL.40)
— BAD-passivated (diagnostis alerts inhibited)
AMP Module
Failure2 (AL.02)
— BAD-maintenance alarm, more diagnosis available
Calibration Due Fault GOOD-
maintenance
demanded
GOOD-
maintenance
demanded
—
GOOD-
maintenance
demanded
GOOD-
maintenance
demanded
GOOD-
maintenance
demanded
UNCERTAIN-
initial value
Outputs in
HOLD
Fault
—
—
—
—
—
—
mA output burn
high
Fault
—
—
—
—
—
—
mA output burn
low
Fault
—
—
—
—
—
—
Error in mA table Fault
—
—
—
—
—
—
mA calculation
error
Fault
—
—
—
—
—
—
mA configuration
error
Fault BAD-maintenance alarm, more diagnosis available
Internal com.
error
Fault BAD-maintenance alarm, more diagnosis available
Chksum err.
(CPU AS)
Fault BAD-maintenance alarm, more diagnosis available
EEPROM err.
(CPU AS)
Fault BAD-maintenance alarm, more diagnosis available
Conductivity too
high
Fault BAD –
maintenance
alarm, more
diagnosis
available
BAD –
maintenance
alarm, more
diagnosis
available
—
BAD –
maintenance
alarm, more
diagnosis
available
BAD –
maintenance
alarm, more
diagnosis
available
BAD –
maintenance
alarm, more
diagnosis
available
Conductivity too
low
Fault BAD –
maintenance
alarm, more
diagnosis
available
BAD –
maintenance
alarm, more
diagnosis
available
—
BAD –
maintenance
alarm, more
diagnosis
available
BAD –
maintenance
alarm, more
diagnosis
available
BAD –
maintenance
alarm, more
diagnosis
available
Temperature too
high
Fault BAD-maintenance alarm, more diagnosis available
Temperature too
low
Fault BAD-maintenance alarm, more diagnosis available
Measurement
unstable
Fault BAD-maintenance alarm, more diagnosis available
Polarization
detected
Fault BAD –
maintenance
alarm, more
diagnosis
available
BAD –
maintenance
alarm, more
diagnosis
available
—
BAD –
maintenance
alarm, more
diagnosis
available
BAD –
maintenance
alarm, more
diagnosis
available
BAD –
maintenance
alarm, more
diagnosis
available
USP limit
exceeded
Fault UNCERTAIN
- process
related, no
maintenance
UNCERTAIN
- process
related, no
maintenance
—
UNCERTAIN
- process
related, no
maintenance
UNCERTAIN
- process
related, no
maintenance
UNCERTAIN
- process
related, no
maintenance
USP margin
exceeded
Fault UNCERTAIN
- process
related, no
maintenance
UNCERTAIN
- process
related, no
maintenance
—
UNCERTAIN
- process
related, no
maintenance
UNCERTAIN
- process
related, no
maintenance
UNCERTAIN
- process
related, no
maintenance