5-168
IM 701310-17E
:TRIGger:EINTerval:EVENt<x>:I2CBus:
NAIGnore:HSMode
Function Sets whether to ignore NON ACK in high speed mode
of the I
2
C bus trigger or queries the current setting.
Syntax
:TRIGger:EINTerval:EVENt<x>:I2CBus:
NAIGnore:HSMode {<Boolean>}
:TRIGger:EINTerval:EVENt<x>:I2CBus:
NAIGnore:HSMode?
<x> = 1 or 2
Example
:TRIGGER:EINTERVAL:EVENT1:I2CBUS:
NAIGNORE:HSMODE ON
:TRIGGER:EINTERVAL:EVENT1:I2CBUS:
NAIGNORE:HSMODE? -> :TRIGGER:EINTERVAL:
EVENT1:I2CBUS:NAIGNORE:HSMODE 1
:TRIGger:EINTerval:EVENt<x>:I2CBus:
NAIGnore:RACCess
Function Sets whether to ignore NON ACK in read access
mode of the I
2
C bus trigger or queries the current
setting.
Syntax
:TRIGger:EINTerval:EVENt<x>:I2CBus:
NAIGnore:RACCess {<Boolean>}
:TRIGger:EINTerval:EVENt<x>:I2CBus:
NAIGnore:RACCess?
<x> = 1 or 2
Example
:TRIGGER:EINTERVAL:EVENT1:I2CBUS:
NAIGNORE:RACCESS ON
:TRIGGER:EINTERVAL:EVENT1:I2CBUS:
NAIGNORE:RACCESS? -> :TRIGGER:EINTERVAL:
EVENT1:I2CBUS:NAIGNORE:RACCESS 1
:TRIGger:EINTerval:EVENt<x>:I2CBus:
NAIGnore:SBYTe (Start Byte)
Function Sets whether to ignore NON ACK in the start byte of
the I
2
C bus trigger or queries the current setting.
Syntax
:TRIGger:EINTerval:EVENt<x>:I2CBus:
NAIGnore:SBYTe {<Boolean>}
:TRIGger:EINTerval:EVENt<x>:I2CBus:
NAIGnore:SBYTe?
<x> = 1 or 2
Example
:TRIGGER:EINTERVAL:EVENT1:I2CBUS:
NAIGNORE:SBYTE ON
:TRIGGER:EINTERVAL:EVENT1:I2CBUS:
NAIGNORE:SBYTE? -> :TRIGGER:EINTERVAL:
EVENT1:I2CBUS:NAIGNORE:SBYTE 1
:TRIGger:EINTerval:EVENt<x>:I2CBus:
SBHSmode?
Function Queries all settings related to the start byte and high
speed mode of the I
2
C bus trigger.
Syntax
:TRIGger:EINTerval:EVENt<x>:I2CBus:
SBHSmode?
<x> = 1 or 2
Example
:TRIGGER:EINTERVAL:EVENT1:I2CBUS:
SBHSMODE? -> :TRIGGER:EINTERVAL:EVENT1:
I2CBUS:SBHSMODE:TYPE HSMODE
:TRIGger:EINTerval:EVENt<x>:I2CBus:
SBHSmode:TYPE
Function Sets the type of the start byte or high speed mode of
the I
2
C bus trigger or queries the current setting.
Syntax
:TRIGger:EINTerval:EVENt<x>:I2CBus:
SBHSmode:TYPE {HSMode|SBYTe}
:TRIGger:EINTerval:EVENt<x>:I2CBus:
SBHSmode:TYPE?
<x> = 1 or 2
Example
:TRIGGER:EINTERVAL:EVENT1:I2CBUS:
SBHSMODE:TYPE HSMODE
:TRIGGER:EINTERVAL:EVENT1:I2CBUS:
SBHSMODE:TYPE? -> :TRIGGER:EINTERVAL:
EVENT1:I2CBUS:SBHSMODE:TYPE HSMODE
:TRIGger:EINTerval:EVENt<x>:LINBus?
Function Queries all settings related to LIN bus signal triggers
of each event.
Syntax
:TRIGger:EINTerval:EVENt<x>:LINBus?
<x> = 1 or 2
Example
:TRIGGER:EINTERVAL:EVENT1:LINBUS?
-> :TRIGGER:EINTERVAL:EVENT1:LINBUS:
BRATE 19200;SOURCE 1
:TRIGger:EINTerval:EVENt<x>:LINBus:
BRATe
Function Sets the LIN bus signal trigger bitrate (data transfer
rate) or queries the current setting.
Syntax
:TRIGger:EINTerval:EVENt<x>:LINBus:
BRATe {<NRf>|USER,<NRf>}
:TRIGger:EINTerval:EVENt<x>:LINBus:
BRATe?
<x> = 1 or 2
<NRf> = 1200, 2400, 4800, 9600, 19200
<NRf> for USER = See the main unit User’s Manual.
Example
:TRIGGER:EINTERVAL:EVENT1:LINBUS:
BRATE 19200
:TRIGGER:EINTERVAL:EVENT1:LINBUS:BRATE?
-> :TRIGGER:EINTERVAL:EVENT1:LINBUS:
BRATE 19200
5.29 TRIGger Group