5-197
IM 701310-17E
Commands
1
2
3
4
5
6
App
Index
:TRIGger:SOURce?
Function Queries all settings related to the trigger source.
Syntax
:TRIGger:SOURce?
Example
:TRIGGER:SOURCE? -> :TRIGGER:SOURCE:
CHANNEL1:COUPLING DC;HFREJECTION OFF;
HYSTERESIS HIGH;LEVEL 1.000E+00;
STATE HIGH;WIDTH 1.000E+00;WINDOW 0;:
TRIGGER:SOURCE:CHANNEL2:COUPLING DC;
HFREJECTION OFF;HYSTERESIS HIGH;
LEVEL 1.000E+00;STATE HIGH;
WIDTH 1.000E+00;WINDOW 0;:TRIGGER:
SOURCE:CHANNEL3:COUPLING DC;
HFREJECTION OFF;HYSTERESIS HIGH;
LEVEL 1.000E+00;STATE HIGH;
WIDTH 1.000E+00;WINDOW 0;:TRIGGER:
SOURCE:CHANNEL4:COUPLING DC;
HFREJECTION OFF;HYSTERESIS HIGH;
LEVEL 1.000E+00;STATE HIGH;
WIDTH 1.000E+00;WINDOW 0;:TRIGGER:
SOURCE:EXTERNAL:LEVEL 0.000E+00;
PROBE 1;:TRIGGER:SOURCE:LOGIC AND
:TRIGger:SOURce:CHANnel<x>?
Function Queries all settings related to the channel of the
trigger source.
Syntax
:TRIGger:SOURce:CHANnel<x>?
<x> = 1 to 4
Example
:TRIGGER:SOURCE:CHANNEL1? -> :TRIGGER:
SOURCE:CHANNEL1:COUPLING DC;
HFREJECTION OFF;HYSTERESIS HIGH;
LEVEL 1.000E+00;STATE HIGH;
WIDTH 1.000E+00;WINDOW 0
:TRIGger:SOURce:CHANnel<x>:COUPling
Function Sets the trigger coupling of the channel or queries the
current setting.
Syntax
:TRIGger:SOURce:CHANnel<x>:
COUPling {AC|DC}
:TRIGger:SOURce:CHANnel<x>:COUPling?
<x> = 1 to 4
Example
:TRIGGER:SOURCE:CHANNEL1:COUPLING AC
:TRIGGER:SOURCE:CHANNEL1:COUPLING?
-> :TRIGGER:SOURCE:CHANNEL1:COUPLING DC
:TRIGger:SOURce:CHANnel<x>:
HFRejection (HighFrequencyREJECTION)
Function Sets the low pass filter (HF rejection) of the channel
or queries the current setting.
Syntax
:TRIGger:SOURce:CHANnel<x>:
HFRejection {<Frequency>|OFF}
:TRIGger:SOURce:CHANnel<x>:HFRejection?
<x> = 1 to 4
<Frequency> = 20MHz or 15kHz
Example
:TRIGGER:SOURCE:CHANNEL1:
HFREJECTION OFF
:TRIGGER:SOURCE:CHANNEL1:HFREJECTION?
-> :TRIGGER:SOURCE:CHANNEL1:
HFREJECTION OFF
Description This command is invalid when the trigger source is
{EXTernal|LINE}.
:TRIGger:SOURce:CHANnel<x>:HYSTeresis
Function Sets the hysteresis of the channel or queries the
current setting.
Syntax
:TRIGger:SOURce:CHANnel<x>:
HYSTeresis {HIGH|LOW}
:TRIGger:SOURce:CHANnel<x>:HYSTeresis?
<x> = 1 to 4
Example
:TRIGGER:SOURCE:CHANNEL1:
HYSTERESIS HIGH
:TRIGGER:SOURCE:CHANNEL1:HYSTERESIS?
-> :TRIGGER:SOURCE:CHANNEL1:
HYSTERESIS HIGH
:TRIGger:SOURce:CHANnel<x>:LEVel
Function Sets the trigger level of the channel or queries the
current setting.
Syntax
:TRIGger:SOURce:CHANnel<x>:
LEVel {<Voltage>|<Current>}
:TRIGger:SOURce:CHANnel<x>:LEVel?
<x> = 1 to 4
<Voltage> and <Current> = See the DL9000 User’s
Manual.
Example
:TRIGGER:SOURCE:CHANNEL1:LEVEL 1V
:TRIGGER:SOURCE:CHANNEL1:LEVEL?
-> :TRIGGER:SOURCE:CHANNEL1:
LEVEL 1.000E+00
5.29 TRIGger Group