5-164
IM 701310-17E
:TRIGger:EINTerval:EVENt<x>:I2CBus:
ADATa:BIT7APsub:ADDRess?
Function Queries all settings related to the 7-bit address of the
7-bit + Sub address of the I
2
C bus trigger.
Syntax
:TRIGger:EINTerval:EVENt<x>:I2CBus:
ADATa:BIT7APsub:ADDRess?
<x> = 1 or 2
Example
:TRIGGER:EINTERVAL:EVENT1:I2CBUS:ADATA:
BIT7APSUB:ADDRESS? -> :TRIGGER:
EINTERVAL:EVENT1:I2CBUS:ADATA:
BIT7APSUB:ADDRESS:PATTERN "10101011"
:TRIGger:EINTerval:EVENt<x>:I2CBus:
ADATa:BIT7APsub:ADDRess:HEXA
Function Sets the 7-bit address of the 7-bit + Sub address of
the I
2
C bus trigger in hexadecimal notation.
Syntax
:TRIGger:EINTerval:EVENt<x>:I2CBus:
ADATa:BIT7APsub:ADDRess:HEXA {<String>}
<x> = 1 or 2
<String> = 2 characters by combining ‘0’ to ‘F’ and ‘X’
(bit 0 is the R/W bit)
Example
:TRIGGER:EINTERVAL:EVENT1:I2CBUS:
ADATA:BIT7APSUB:ADDRESS:HEXA "AB"
:TRIGger:EINTerval:EVENt<x>:I2CBus:
ADATa:BIT7APsub:ADDRess:PATTern
Function Sets the 7-bit address of the 7-bit + Sub address of
the I2C bus trigger in binary notation or queries the
current setting.
Syntax
:TRIGger:EINTerval:EVENt<x>:I2CBus:
ADATa:BIT7APsub:ADDRess:PATTern
{<String>}
:TRIGger:EINTerval:EVENt<x>:I2CBus:
ADATa:BIT7APsub:ADDRess:PATTern?
<x> = 1 or 2
<String> = 8 characters by combining ‘0’, ‘1’, and ‘X’
(bit 0 is the R/W bit)
Example
:TRIGGER:EINTERVAL:EVENT1:I2CBUS:ADATA:
BIT7APSUB:ADDRESS:PATTERN "10101011"
:TRIGGER:EINTERVAL:EVENT1:I2CBUS:ADATA:
BIT7APSUB:ADDRESS:PATTERN? -> :TRIGGER:
EINTERVAL:EVENT1:I2CBUS:ADATA:
BIT7APSUB:ADDRESS:PATTERN "10101011"
:TRIGger:EINTerval:EVENt<x>:I2CBus:
ADATa:BIT7APsub:SADDress?
Function Queries all settings related to the Sub address of the
7-bit + Sub address of the I
2
C bus trigger.
Syntax
:TRIGger:EINTerval:EVENt<x>:I2CBus:
ADATa:BIT7APsub:SADDress?
<x> = 1 or 2
Example
:TRIGGER:EINTERVAL:EVENT1:I2CBUS:ADATA:
BIT7APSUB:SADDRESS? -> :TRIGGER:
EINTERVAL:EVENT1:I2CBUS:ADATA:
BIT7APSUB:SADDRESS:PATTERN "10101011"
:TRIGger:EINTerval:EVENt<x>:I2CBus:
ADATa:BIT7APsub:SADDress:HEXA
Function Sets the Sub address of the 7-bit + Sub address of
the I
2
C bus trigger in hexadecimal notation.
Syntax
:TRIGger:EINTerval:EVENt<x>:I2CBus:
ADATa:BIT7APsub:SADDress:HEXA
{<String>}
<x> = 1 or 2
<String> = 2 characters by combining ‘0’ to ‘F’ and ‘X’
Example
:TRIGGER:EINTERVAL:EVENT1:I2CBUS:ADATA:
BIT7APSUB:SADDRESS:HEXA "EF"
:TRIGger:EINTerval:EVENt<x>:I2CBus:
ADATa:BIT7APsub:SADDress:PATTern
Function Sets the Sub address of the 7-bit + Sub address of
the I
2
C bus trigger in binary notation or queries the
current setting.
Syntax
:TRIGger:EINTerval:EVENt<x>:I2CBus:
ADATa:BIT7APsub:SADDress:PATTern
{<String>}
:TRIGger:EINTerval:EVENt<x>:I2CBus:
ADATa:BIT7APsub:SADDress:PATTern?
<x> = 1 or 2
<String> = 8 characters by combining ‘0,’ ‘1,’ and ‘X’
Example
:TRIGGER:EINTERVAL:EVENT1:I2CBUS:ADATA:
BIT7APSUB:SADDRESS:PATTERN "10101011"
:TRIGGER:EINTERVAL:EVENT1:I2CBUS:ADATA:
BIT7APSUB:SADDRESS:PATTERN?
-> :TRIGGER:EINTERVAL:EVENT1:I2CBUS:
ADATA:BIT7APSUB:SADDRESS:
PATTERN "10101011"
:TRIGger:EINTerval:EVENt<x>:I2CBus:
ADATa:TYPE
Function Sets the address type of the I
2
C bus trigger or queries
the current setting.
Syntax
:TRIGger:EINTerval:EVENt<x>:I2CBus:
ADATa:TYPE {BIT10address|BIT7ADdress|
BIT7APsub}
:TRIGger:EINTerval:EVENt<x>:I2CBus:
ADATa:TYPE?
<x> = 1 or 2
Example
:TRIGGER:EINTERVAL:EVENT1:I2CBUS:
ADATA:TYPE BIT10ADDRESS
:TRIGGER:EINTERVAL:EVENT1:I2CBUS:ADATA:
TYPE? -> :TRIGGER:EINTERVAL:EVENT1:
I2CBUS:ADATA:TYPE BIT10ADDRESS
5.29 TRIGger Group