5-63
IM 701310-17E
Commands
1
2
3
4
5
6
App
Index
:ANALysis:XY<x>:MEASure:XYINteg:
POLarity
Function Sets the integration polarity of the XY display or
queries the current setting.
Syntax
:ANALysis:XY<x>:MEASure:XYINteg:
POLarity {CCW|CW}
:ANALysis:XY<x>:MEASure:XYINteg:
POLarity?
<x> = 1 or 2
Example
:ANALYSIS:XY1:MEASURE:XYINTEG:
POLARITY CCW
:ANALYSIS:XY1:MEASURE:XYINTEG:POLARITY?
-> :ANALYSIS:XY1:MEASURE:XYINTEG:
POLARITY CCW
:ANALysis:XY<x>:MEASure:XYINteg:
VALue?
Function Queries the integral value of the XY display.
Syntax
:ANALysis:XY<x>:MEASure:XYINteg:VALue?
<x> = 1 or 2
Example
:ANALYSIS:XY1:MEASURE:XYINTEG:VALUE?
-> :ANALYSIS:XY1:MEASURE:XYINTEG:
VALUE 1.000E+00
:ANALysis:XY<x>:TRANge (Time Range)
Function Sets the measurement range of the XY display or
queries the current setting.
Syntax
:ANALysis:XY<x>:TRANge {<NRf>,<NRf>}
:ANALysis:XY<x>:TRANge?
<x> = 1 or 2
<NRf> = –5 to 5 div
:ANALysis:XY<x>:WINDow
Function Sets the measurement source window of the XY
display or queries the current setting.
Syntax
:ANALysis:XY<x>:WINDow {MAIN|Z1|Z2}
:ANALysis:XY<x>:WINDow?
<x> = 1 or 2
Example
:ANALYSIS:XY1:WINDOW MAIN
:ANALYSIS:XY1:WINDOW? -> :ANALYSIS:XY1:
WINDOW MAIN
:ANALysis:XY<x>:MEASure:CURSor:Y<x>:
VALue?
Function Queries the voltage value at the vertical cursor of the
XY display.
Syntax
:ANALysis:XY<x>:MEASure:CURSor:Y<x>:
VALue?
<x> of XY<x> = 1 or 2
<x> of Y<x> = 1 or 2
Example
:ANALYSIS:XY1:MEASURE:CURSOR:Y1:VALUE?
-> :ANALYSIS:XY1:MEASURE:CURSOR:Y1:
VALUE 1.000E+00
:ANALysis:XY<x>:MEASure:MODE
Function Sets the automated measurement mode of the XY
display or queries the current setting.
Syntax
:ANALysis:XY<x>:MEASure:MODE {CURSor|
OFF|XYINteg}
:ANALysis:XY<x>:MEASure:MODE?
<x> = 1 or 2
Example
:ANALYSIS:XY1:MEASURE:MODE CURSOR
:ANALYSIS:XY1:MEASURE:MODE?
-> :ANALYSIS:XY1:MEASURE:MODE CURSOR
:ANALysis:XY<x>:MEASure:XYINteg?
Function Queries all settings related to the integration of the
XY display.
Syntax
:ANALysis:XY<x>:MEASure:XYINteg?
<x> = 1 or 2
Example
:ANALYSIS:XY1:MEASURE:XYINTEG?
-> :ANALYSIS:XY1:MEASURE:XYINTEG:
LOOP CLOSE;POLARITY CCW
:ANALysis:XY<x>:MEASure:XYINteg:LOOP
Function Sets the integration mode of the XY display or queries
the current setting.
Syntax
:ANALysis:XY<x>:MEASure:XYINteg:LOOP
{CLOSe|OPEN}
:ANALysis:XY<x>:MEASure:XYINteg:LOOP?
<x> = 1 or 2
Example
:ANALYSIS:XY1:MEASURE:XYINTEG:
LOOP CLOSE
:ANALYSIS:XY1:MEASURE:XYINTEG:LOOP?
-> :ANALYSIS:XY1:MEASURE:XYINTEG:
LOOP CLOSE
5.3 ANALysis Group