Chapter 6 How to use the Multi-functions
39
6-5. Counting
function
(Note) The counting function cannot be used for transactions or certifications stipulated by the Measurement
Act.
* When "NOT LEGAL" is displayed in the lower left of the display, you
cannot use the results for transactions or certifications stipulated by the
Measurement Act.
Counting function
Application Counting
work
Weighing method
The weight per item (unit weight) is measured from a specified quantity of items
to calculate the number of items from the weight data of the whole items.
Advantage
This function is suitable for quantity counting work.
Setting values
• Number of weighing samples; and weight measurement (once)
• Lower limit quantity
• Upper limit quantity (when the upper/lower limit function is used)
Indication on Main-
display
Number of items
Indication on Sub-
display
Measured weight
Max number of
program entries
99 programs
For counting work, enter the number of samples and then measure the weight of the samples.
(Enter the number of samples)
(The samples are weighed)
Place the specified number of
the samples.
* The weight of the container is
subtracted as tare.
(Note) If the measured weight of samples is less than 4 scale intervals, the system does not calculate the unit
weight. If the unit weight is less than 1 scale interval, a weighing error may occur. Therefore, it is
recommended to use the counting function for items weighing at least 1 scale interval.
* When the lower limit quantity is set, the lower limit
quantity is displayed in the Sub-display at the lower
right instead of weight.
Quantity
Measured
Weight