XO FLEX – Instructions for use
Infection
control and maintenance checklist
77
What
How
Parts/consumables needed
•
Handles on instrument
bridge, light and unit HD
Display
•
Cuspidor bowl, gold trap and
protection disk
•
Cup and cup holder
Clean and disinfect all surfaces.
See section 5.5
See section 10.3.
Every evening
Clean and disinfect the suction
lines
See section 5.19
XO Suction Disinfection cartridges
See section 10.3.
Clean suction filters
See section 5.19.2
XO Silicone Grease, O-rings
See section 10.3.
Start the overnight water line
disinfection procedure (step 1).
See section 5.20.1
Switch the unit off, leaving the
unit instruments in the
instrument holder on the
cuspidor.
XO Water Clean cartridges
See section 10.3.
When the unit has not been used for several days – after a long weekend, holidays etc.
Start the intensive water line
disinfection procedure.
See section 5.20.2
XO Water Clean cartridges
See section 10.3.
Every month
Control the water disinfection
system
See section 6.7
Test strip
See section 10.3.
Empty coarse filter in cuspidor
valve, if installed.
See section 6.8
Measure curing depth of XO
ODONTOCURE
See section 6.2
XO ODONTOCURE testing device
– see section 10.3 – and/or LED UV
measuring device
Every year
12, 36 etc. months after
installation of the unit: Preventive
service and safety inspection A.
See section 8.2.1
XO FLEX Service Kit A
See section 10.3.
24, 48 etc. months after
installation of the unit: Preventive
service and safety inspection B.
See section 8.2.2
XO FLEX Service Kit B
See section 10.3.
Replace:
•
suction hoses
•
gold trap in cuspidor
•
syringe tip
XO FLEX Infection Control Kit
See section 10.3.