XO FLEX – Instructions for use
Cleaning
and infection control
54
When cleaning the surfaces – do not use an excessive amount of liquid – and make
sure that the liquid has evaporated before fitting instrument holder pad, cuspidor,
cuspidor protection disc and cup holder.
5.4
CLEANING OF COMFORT AND SKAI FABRIC
Wipe the fabric with a dry or moisturized lint-free piece of soft cloth after each patient.
Apply Dürr FD 360 to a dry lint-free cloth to remove stains as soon as possible.
After the last patient wipe the fabric using a lint-free cloth with Dürr FD 360. Wipe excess of Dürr
FD 360 with a dry cloth.
Please note: Dürr FD 360 shall be applied to the cloth – never directly to the fabric!
5.5
GENERAL DISINFECTION OF UNIT SURFACES
Disinfect the surfaces of the unit using a lint-free cloth with Dürr FD 333 or Dürr 366 – please see
details in section 10.3.
Look out for disinfection liquid dripping from the cloth.
When disinfecting surfaces – do not use an excessive amount of disinfectant – and
make sure that the disinfectant has evaporated before fitting instrument holder pad,
cuspidor, cuspidor protection disc and cup holder.
5.6
INFECTION CONTROL INSTRUMENT AND BRIDGE PROTECTION PAD
The instrument pad and bridge protection pad can be cleaned in a thermo-disinfector and
autoclaved.
Figure 67 – Instrument pad and bridge protection pad
5.7
AUTOCLAVING HANDLES
The instrument bridge, light and suction hose holder handles can be removed to be autoclaved by
pressing the knob (1) at the end of the handle – see Figure 68.