XO FLEX – Instructions for use
Cleaning
and infection control
57
Figure 71 – Sealing plug in XO OSSEO motor shaft
Do not leave the motor and hose in the autoclave overnight. Doing so may cause
damages to the motor.
Do not wash the motor and the hose in the thermo-disinfector. Doing so may cause
damages to the motor.
Do not lubricate through the open front end of the motor.
5.12
INFECTION CONTROL XO ODONTOSON 360
The instruments can be cleaned and sterilized using all clinically acceptable methods.
The handpiece can be cleaned and disinfected with Dürr FD 333 or autoclaved (max 121°C).
5.13
INFECTION CONTROL XO ODONTOCURE CURING LIGHT
Use the supplied cross infection protection sleeves to decrease the risk of cross contamination
and prevent the composite from bonding to the end of the rod.
Every day the light rod should be examined for stuck restorative material and mechanical
damages.
Remove the light rod from the instrument by pulling it with your hand.
Autoclave the rod and protection cap.
Disinfect the handpiece and hose with Dürr FD 333.