XO FLEX – Instructions for use
Configuration
49
Figure 66 – Configuration using the foot control
Table 5 – Unit and patient chair parameters that can be configured
Parameter
Display text
Factory default
configuration
Possible settings
Enabling XO Peristaltic
Pump
PUMP
None
None, POS. 1 – 6
Cuspidor flush stops
after
FLUSH/S
15 s
10 – 600
Cup filler stops after
CUP/S
10 s
5 – 15 s
Cuspidor flush starts
automatically when chair
reaches entry position
FLUSH/CH
Enabled
Disabled, enabled
Cuspidor flush starts
automatically when cup
filler stops
FLUSH/CF
Enabled
Disabled, enabled
Light on when patient
chair reaches a working
position or previous
position.
Light off when chair
moves towards entry, 0-
position or working
positions 1 or 2.
LAMP/CH
Disabled
Disabled, enabled
Light on/off with foot
control
LAMP/PD
Disabled
Disabled, enabled
Reset unit and patient
chair functions to
“factory standard”
RESET
-
Yes, No
Rinse position = entry
position (0-position)
RINSE = 0
Disabled
Disabled, enabled
4.3
CONFIGURATION OF UNIT INSTRUMENTS
1.
Lift the instrument to be programmed forward
2.
Activate the configuration button (1) until the display shows the instrument setting – see
Figure 64
3.
Activate
▲
/
▼
or
⬇
– see Figure 66 – to browse the parameter to change (see Table 6)