XO FLEX – Instructions for use
Cleaning
and infection control
56
Wipe the cloth from the instrument coupling towards the plug to prevent disinfection liquid from
leaking into the instrument coupling! See Figure 70.
Figure 70 – Wiping direction, instrument hose-cleaning
Instrument suspensions (not the instrument hoses) can be washed in a thermo-disinfector.
Do not wash instrument hoses in a thermo-disinfector!
When refitting the instrument hoses, the unit will inform you in case an instrument hose is fitted to
a “wrong” position.
Make sure that the instrument hose plugs and sockets in the instrument bridge are
completely dry before the hoses are re-mounted on the instrument bridge!
5.9
INFECTION CONTROL LUZZANI SYRINGE
Follow the instructions for use supplied by Luzzani.
5.10
INFECTION CONTROL BIEN-AIR MICROMOTORS
Follow the instructions for use supplied by Bien-Air.
5.11
AUTOCLAVING OF XO OSSEO MOTOR
The recommended infection control procedure is to autoclave the motor with hose after each use.
Prior to autoclaving push the end of the sealing plug into the open center of the
motor shaft – see Figure 71.
Failure to autoclave the motor with the sealing plug in place may cause damages to
the motor.