XO FLEX – Instructions for use
Unit
messages and remedial actions
73
7.2
SERVICE MESSAGES AND REMEDIAL ACTIONS
Below you see an overview of the service messages that may appear on the instrument bridge
display.
Table 9 – Service messages
Service message
Reason for message
Remedial action
CHECK YELLOW BOTTLE
The unit will soon run out of XO
Suction Disinfection.
Make sure you have XO Suction
Disinfection cartridges (AN-354)
in stock.
CHECK WHITE BOTTLE
The unit will soon run out of XO
Water Clean
Make sure you have XO Water
Clean cartridges (AO-980) in
stock.
FLUSH MM MIN
Water lines are being flushed. MM
= remaining minutes to completion
of process.
Please wait – do not shut off
power or activate any buttons.
FLUSH OK
Water line flushing completed.
Unit operational again.
NEXT SERVICE
YYYY-MM-DD
Less than 30 days to next
preventive service and safety
inspection.
Call your XO authorized service
provider and require a preventive
service and safety inspection.
SERVICE OVERDUE
YYYY-MM-DD
Date for preventive service and
safety inspection is overdue.
Call your XO authorized service
provider and require a preventive
service and safety inspection.
SUCTION OK
Suction disinfection process
completed. Unit operational again.
SUCTION X %
The suction disinfection process is
ongoing. X = remaining time in %.
Please wait – do not shut off
power or activate any buttons.
WATER CLEAN ACTIVE
WAIT MM
Water tank discharging water. MM
= remaining minutes to completion
of process.
Please wait – do not shut off
power or activate any buttons.
PUMP NOT READY
Pump detached or no connection
to unit.
Attach pump correctly.
NO PUMP
Pump not connected to the
selected suspension/instrument
Attach pump correctly. Check
configuration of the pump.
TOO HOT
If XO ODONTOCURE is activated
continuously the message “TOO
HOT” is displayed and the
instrument is turned off.
Depending on the temperature,
the light is ready for re-activation
after a few seconds.