XO FLEX – Instructions for use
Operation
21
3.8
INSTRUMENT CONTROL
FOOT CONTROL OF INSTRUMENTS
You activate the first instrument lifted forward with the foot control pedal. To avoid unintentional
activation of a second instrument lifted forward you will only be able to activate the second
instrument when the first instrument is laid back and the foot control has been released.
The syringe may be used even though another instrument is lifted forward.
The intraoral video camera may be used simultaneously with another activated instrument – but
control of camera functions via the foot control is only possible when the video camera is the only
instrument lifted forward.
INSTRUMENTS ON THE INSTRUMENT BRIDGE
Up to 6 instruments may be fitted to the instrument bridge.
Figure 18 – Numbering of unit instruments
The instruments are numbered 1 – 6 starting from the left.
Your XO authorized service provider may add instruments or change the order of the instruments –
the multifunction syringe, however, must always be placed to the left (right-handed operator) or to
the right (left-handed operator).
INSTRUMENT BRIDGE DISPLAY
Data concerning the selected instrument is shown on the instrument bridge display – see Figure 19
– and the instrument may be controlled using the foot control.
Figure 19 – Instrument bridge display