XO FLEX – Instructions for use
Cleaning
and infection control
60
The cuspidor bowl, the gold trap, the cuspidor protection disc, and the cup holder can be
disinfected in a thermo-disinfector.
5.19
DISINFECTION OF SUCTION LINES
GENERAL
The suction valves and tubes inside the unit are constantly flushed with clean water while the
suction is activated.
To prevent the suction system from clogging and breaking down, it is also necessary to clean and
disinfect the suction system daily.
1.
Lift both suction hoses from the suction holder
2.
Remove the covers from both suction nozzles – see Figure 75
3.
Activate the suction disinfection button (2) – see Figure 76 – to deactivate the suction
4.
Place the two suction nozzles on the suction hose connectors (3) and activate the button (2)
again – see Figure 76 – hereafter the flushing procedure starts
Figure 75 – Removal of suction nozzle covers
Figure 76 – XO Suction Disinfection (1) XO Suction Disinfection cartridge
(2) button for activation of XO Suction Disinfection (3) suction hose connectors
The process takes approximately 6 minutes, depending on the suction power.
The display shows the progress of the disinfection process.
When “SUCTION OK” appears on the display, the process has been carried out successfully.