XO FLEX – Instructions for use
Configuration
50
4.
Activate
►
to select the parameter to change
5.
Use
▲
/
▼
to change the value of the selected parameter or
⬇
to increase the value
6.
Activate
◄
to save the new parameter
7.
If an additional instrument parameter is to be configured, use
▲
/
▼
until the display shows
the parameter in question
8.
Place the instrument on the bridge again
Table 6 – Unit instrument parameters that can be configured
Parameter
Display text
Factory default
configuration
Possible settings
Bien-Air MC3 / MX2 micro motor
Maximum speed
–
Level 3
LEVEL 3
40,000 RPM
100 – 40,000 + 10% RPM
Maximum speed – Level
2
LEVEL 2
10,000 RPM
100 – 40,000 + 10% RPM
Maximum speed – Level
1
LEVEL 1
4.000 RPM
100 – 40,000 + 10% RPM
Amount of spray water –
Level 3
W LEVEL 3
Without contra-
angle (what you see
on the display): 100
ml/min
With contra-angle:
approximately 70
ml/min
Without contra-angle: 10 –
100 ml/min
Amount of spray water –
Level 2
W LEVEL 2
Without contra-
angle (what you see
on the display): 50
ml/min
With contra-angle:
approximately 50
ml/min
Without contra-angle: 10 –
100 ml/min
Amount of spray water –
Level 1
W LEVEL 1
Without contra-
angle (what you see
on the display): 35
ml/min
With contra-angle:
approximately 35
ml/min
Without contra-angle: 10 –
100 ml/min
Amount of spray air
AIR
70%
0 – 100%
Automatic chip blow
CHIP
Enabled
Enabled, Disabled
Spray selection
4 STATES
Disabled
Enabled, Disabled
XO OSSEO motor
Maximum speed
–
Level 3
LEVEL 3
40,000 RPM
100 – 40,000 + 10% RPM
Maximum speed – Level
2
LEVEL 2
16,000 RPM
100 – 40,000 + 10% RPM
Maximum speed – Level
1
LEVEL 1
300 RPM
100 – 40,000 + 10% RPM
4
Please note that it is the speed of the micromotor itself. When you fit a contra-angle the speed of the bur may be changed!
5
Please note that it is the speed of the micromotor itself. When you fit a contra-angle the speed of the bur may be changed!