XO FLEX – Instructions for use
Configuration
51
Parameter
Display text
Factory default
configuration
Possible settings
Amount of spray water –
Level 3
W LEVEL 3
50 ml/min
Peristaltic pump i
rrigation
flow rate:
10 – 90 ml/min ±
20%
Amount of spray water –
Level 2
W LEVEL 2
30 ml/min
Peristaltic pump i
rrigation
flow rate:
10 – 90 ml/min ±
20%
Amount of spray water –
Level 1
W LEVEL 1
10 ml/min
Peristaltic pump i
rrigation
flow rate:
10 – 90 ml/min ±
20%
Turbine (Air instrument)
Drive air – Level 3
LEVEL 3
100%
0 – 100%
Drive air – Level 2
LEVEL 2
90%
0 – 100%
Drive air – Level 1
LEVEL 1
70%
0 – 100%
Amount of spray water –
Level 3
W LEVEL 3
Without turbine
hand-piece (what
you see on the
display): 100 ml/min
With turbine hand-
piece:
approximately
70
ml/min
Without turbine hand-piece:
10 – 100 + 20% ml/min
Amount of spray water –
Level 2
W LEVEL 2
Without turbine
hand-piece (what
you see on the
display): 50 ml/min
With turbine hand-
piece:
approximately
50
ml/min
Without turbine hand-piece:
10 – 100 + 20% ml/min
Amount of spray water –
Level 1
W LEVEL 1
Without turbine
hand-piece (what
you see on the
display): 35 ml/min
With turbine hand-
piece:
approximately
35
ml/min
Without turbine hand-piece:
10 – 100 + 20% ml/min
Amount of spray air
AIR
70%
0 – 100%
Automatic chip blow
CHIP
Enabled
Enabled, Disabled
XO ODONTOSON 360 or other ultrasonic scaler
Power – Level 3
LEVEL 3
100%
0 – 100%
Power – Level 2
LEVEL 2
70%
0 – 100%
Power – Level 1
LEVEL 1
40%
0 – 100%
Amount of irrigation
water – Level 3
W LEVEL 3
40 ml/min
10 – 90 + 10% ml/min
Amount of irrigation
water – Level 2
W LEVEL 2
30 ml/min
10 – 90 + 10% ml/min