XO FLEX – Instructions for use
Cleaning
and infection control
64
4.
When "FLUSH OK" appears on the display, the process has been carried out successfully
5.
Place the unit instruments on the instrument bridge
Please note that manual cancellation of this process is not possible.
In case of emergency you may switch off the unit, place the unit instruments on the instrument
bridge, remove the water disinfection instrument holder etc. – and switch on the unit again.
Please note that XO Water Clean and XO Suction Disinfection cannot be activated simultaneously!
INTENSIVE WATER TREATMENT
Intensive water treatment should be carried out when the bacterial load of the unit water lines is
considered higher than usual, for instance, after vacation or other longer periods without usage.
1.
Place all water-bearing unit instruments in the instrument holder on the cuspidor (make sure
that the heating on the 6F syringe is switched off)
2.
Attach the cup filler outlet as shown in Figure 80
3.
Activate the button (2) in Figure 80 and the instrument water lines are flushed
4.
Watch the display for the remaining time of the disinfection process
When "FLUSH OK" appears in the display, the process has been carried out successfully.
Please note that manual cancellation of this process is not possible.
In case of emergency you may switch off the unit, place the unit instruments on the instrument
bridge, remove the water disinfection instrument holder etc. – and switch on the unit again.
We recommend that this procedure be performed once after the weekend.
When the unit has not been used for a week or more, we recommend that the
intensive procedure be performed three consecutive times.
REPLACEMENT OF XO WATER CLEAN CARTRIDGE
When the cartridge is almost empty a “warning-sound” is heard and the message “CHECK WHITE
BOTTLE” appears in the unit display.
The XO Water Clean cartridge is exchanged as follows:
1.
Lower the left-hand service panel – Figure 81
2.
Pull the white handle forward to get access to the cartridge
3.
Pull the cartridge downwards to get it free
4.
Replace the cartridge, push the white handle in and close the service panel