MAINTENANCE AND REPAIRS
Danger of electric shock!
Do not attempt to open the product unless you are an authorized service provider!
6.1
FOOT CONTROL
The foot control is fitted with four rubber feet providing a stable attachment to the floor.
If the rubber feet and the floor become greasy with soap the friction may be reduced and the foot
control may slide on the floor when activated.
If necessary, clean the rubber feet and floor with petroleum benzine to avoid the foot
control from sliding when activated!
Figure 82 – Cleaning of foot control “rubber feet”
6.2
XO OSSEO MOTOR
MOTOR REPAIR
If necessary the XO OSSEO motor must be repaired at the factory.
6.3
XO ODONTOSON
HANDPIECE REPAIR
If necessary the XO ODONTOSON handpiece must be repaired at the factory.
TIGHTENING / EXCHANGING THE FERRITE ROD
If an instrument has lost its efficiency, it may help to re-tighten the ferrite rod – see Figure 83 – and
check that the inside of the hand-piece is free of water and foreign bodies.