XO FLEX – Instructions for use
Operation
23
Enable/disable automatic chip blow with
O
.
When the automatic chip blow is enabled a short burst of high-pressure air dries the
preparation each time an instrument with spray water stops.
Using the automatic chip blow function significantly reduces the number of shifts
between rotation instrument and syringe.
Choose between three levels of spray water:
Increase amount of water with
►
.
Decrease amount of water with
◄
.
To avoid necrosis, it is generally recommended to work with a “wet” spray that gives
at least 50 ml spray water per minute measured with the contra-angle fitted!
In some cases – e.g. when preparing a cavity that is not close to the pulp – and you
want to minimize the spray aerosol – it may be acceptable to use less water in the
spray.
Choose between three maximum speed levels:
Increase maximum speed with
▲
.
Decrease maximum speed with
▼
.
3.11
XO OSSEO MOTOR
XO OSSEO motor is intended for bone surgery in general and for dental implantology.
XO OSSEO motor is intended to be used with surgical contra-angles and handpieces with
couplings as specified in ISO 3964, type 1.
SET-UP
Appropriate sterile handling procedures must be observed to ensure and maintain sterile
conditions:
1.
Remove the sterilized motor and hose from the autoclave bag – or wrap motor and hose
immediately before use.
2.
Connect the motor hose to the matching connector at the rear of the instrument bridge –
see section 5.8. The instrument suspension shall be configured as MOTOR.
3.
Prepare the peristaltic pump as described in section 3.16.
4.
Remove the sterilized contra-angle from the autoclave bag just prior to surgery.
5.
Slide the contra-angle firmly in place on the motor shaft.
6.
Attach the sterile irrigation tube to the contra-angle using one of the two supplied outlets –
See Figure 21. The choice of outlet depends on the contra-angle.
Содержание FLEX
Страница 1: ...XO FLEX INSTRUCTIONS FOR USE...
Страница 2: ...XO FLEX Instructions for use 2...
Страница 20: ...XO FLEX Instructions for use Operation 20 Figure 17 Six different patient positions...
Страница 88: ...XO FLEX Instructions for use Symbols 88 Chair base joystick south Chair base joystick east...
Страница 94: ......