XO FLEX – Instructions for use
Maintenance
and repairs
69
In case of faulty performance, the fiber rod may be replaced and retested. In case the
problem persists, technical assistance from an authorized service provider is
required.
6.5
XO PERISTALTIC PUMP
Please refer to Table 9 on page 73 to see an overview of the service messages that may appear on
the display while using XO Peristaltic Pump.
6.6
ADJUSTMENT OF THE HAND INSTRUMENT TABLE
You can adjust the angle of the hand instrument table using a 4 mm Allen key.
Figure 87 – Adjustment of the hand instrument table
6.7
CONTROL OF THE WATER DISINFECTION SYSTEM
If the unit is fitted with a water disinfection system, we recommend that you control the function
every month:
Disinfect the unit water lines as described in 5.20.2
1)
A: Place a clean cup in the cup holder (units with cuspidor and cup filler) and activate the
cup filler
1)
B: Using the syringe fill about 20 ml of water in a clean cup (units without cuspidor and cup
filler)
2)
Dip a peroxide test strip in the water for one second
3)
Shake off excess liquid
4)
Wait 45 seconds
5)
Compare with the color scale below:
Figure 88 – Water quality color scale
The test strip shall indicate a H
2
O value between 150 and 300 mg/l.