Figure 21 Attaching the irrigation tube
OPERATION
Lift the motor forward and control it with the foot control.
Activate the motor clockwise in the range 100 RPM – select maximum speed
(see below) with
⮕
or counterclockwise with
⬅
.
See the maximum speed on the display when the foot control is not activated. See the actual
speed on the display when the motor is running.
Figure 22 – Foot control of XO OSSEO motor
Switch irrigation on/off with
X
.
Choose between three levels of irrigation:
Increase amount of irrigation with
►
.
Decrease amount of irrigation with
◄
.
To avoid necrosis, it is generally recommended to work with an amount of irrigation
of at least 50 ml per minute measured with the contra-angle fitted!
Choose between three maximum speed levels:
Increase maximum speed with
▲.
Decrease maximum speed with
▼.
3
Please note that this is the speed of the motor itself. When you fit a contra-angle the speed of the bur may be changed!