Alphabetical reference
Commands and Queries Reference
OUT
program the state of the Global Output Enable Flag of the XMP 2600.
Syntax
OUT
<int>
Parameters
the desired state of the Global Output Enable. A
value of “0” disables and a value of “1” enables.
Category Output
Type:
Overlapped
OUT
program the state of the Power Module Output.
Syntax
OUT
<ch>,<int>
Parameters
the number of the Power Module to be programmed.
the desired state of the Power Module’s output: a
value of “0” disables and a value of “1” enables.
Category Output
Type:
Overlapped
OUT?
return the current state of the Global Output Enable Flag. The result is
decimal value in the range 0 to 1.
Syntax
OUT?
Parameters none
Category Output
Type:
Sequential
OUT?
return the current state of the Power Module’s output. The result is a
decimal value in the range 0 to 1.
Syntax
OUT?
<ch>
Parameters
the number of the channel to be queried.
Category Output
Type:
Sequential
OVRST
reset OVP and OCP errors reported by the Power Module. Clear all
Warning, Foldback and Fault indications from the Status Structure.
The Power Module’s output is restored to its pre-fault state.
Syntax
OVRST
<ch>
Parameters
the number of the Power Module to operated upon.
Category Protection
Type:
Overlapped
OVSET
program the Power Module’s Over Voltage Protection.
Syntax
OVSET
<ch>,<value>
Parameters
the number of the Power Module to be programmed.
the value to be assigned to the OVP threshold. A
value can only be programmed when PROT mode is
Manual. The upper limit of the Programmed value is
VMAX + 1. The lower limit of the programmed
value is VSET.
Category Protection
Type:
Sequential
54
XMP 2600 Programming Manual
rev. 1.1