Alphabetical reference
Commands and Queries Reference
TOSET
set the remote communication time out period.
Syntax
TOSET
<int>
Parameters
the amount of time to wait between remote
communication events before a time out event is
declared. The allowed values are in the range 1 to
255 and the time units are seconds. A value of “0”
will not change the programmed time out period but
will cause the feature to be disabled (same as
“TOEN 0”).
Category System
Type:
Sequential
TOSET?
return the programmed time out period of the remote communication
time out feature.
Syntax
TOSET?
Parameters none
Category System
Type:
Sequential
TRIG
program the Power Module’s reaction to a software or hardware
trigger.
Syntax
TRIG
<ch>,<int>
Parameters
the number of the Power Module to be programmed.
the action to be performed by the Power Module
when a trigger is received:
0 = set output to values programmed while armed.
1 = toggle output on/off.
2 = initiate ramp to VSET programmed while armed.
3 = initiate ramp to ISET programmed while armed.
4 = output on.
5 = output off.
6 = no operation.
7 = un-arm.
Category
Trigger and Sync Type:
Sequential
TRIG?
return the current reaction to trigger for the Power Module. Result is a
decimal number in the range 0 to 7.
Syntax
TRIG?
<ch>
Parameters
the number of the channel to be queried.
Category
Trigger and Sync Type:
Sequential
TRIGER
send a global trigger (to all Power Modules).
Syntax
TRIGER
Parameters none
Category
Trigger and Sync Type:
Overlapped
64
XMP 2600 Programming Manual
rev. 1.1