Status reporting
Self Test Results
The following illustrates the System Status Structure:
EVENT
OR
&
ENABLE
PRIMARY
OUTPUT
SET
QUEUE
EVENT
OR
&
channels
ENABLE
SRQ
STATUS
GEN
OR
&
&
OR
ENABLE
ENABLE
IST
Self Test Results
The XMP 2600 enables its user to conduct two types of self-tests: Power Module self-test
and main controller’s self-test.
Power
Modules
The Power Module Test or Confidence Test result is made out of two bytes. The first byte
is the result of the latest performed test, and the second byte is the result of an on-going
confidence test of the analog section of the controlling circuits.
The results of the tests can be remotely read using the CONF? <ch> query.
The CONF? <ch> query will return the results of the latest test performed
and initiates a run of the Confidence Test.
XMP 2600 Programming Manual
rev. 1.1
75