Operating Time
3 hours at 5 minutes working mode.
The device enters Sleep mode if idle for 2 minutes.
Once the device is idle for 2 minutes it enters Sleep mode and if it remains idle in Sleep mode for
another 10 minutes the device powers down.
Charging Time
5 hours (through the docking station using a Welch Allyn approved battery)
All specifications are subject to change.
Operating environment
Criterion
Environmental conditions
Temperature
+ 10 °C to + 35 °C
Relative humidity
30 % to 90 %
Atmospheric pressure
800 hPa to 1060 hPa
Storage environment
Criterion
Environmental conditions
Temperature
− 10 °C to + 55 °C
Relative humidity
10 % to 95 %
Atmospheric pressure
700 hPa to 1060 hPa
Transportation environment
Criterion
Environmental conditions
Temperature
− 40 °C to + 70 °C
Relative humidity
10 % to 95 %
Atmospheric pressure
500 hPa to 1060 hPa
Device lifetime
The lifetime of the device is five (5) years. The manufacture date of the device can be
found on the device labels.
Welch Allyn will service RetinaVue 100 Imager devices that are within their lifetime.
Firmware updates and support may require an annual subscription service after the initial
one (1) year warranty period. The expected battery life is at least one (1) year. If the
RetinaVue 100 Imager device fails to hold a charge, a new battery can be ordered.
72 Specifications
Welch Allyn RetinaVue™ 100 Imager
Содержание RetinaVue 100 Imager
Страница 1: ...Welch Allyn RetinaVue 100 Imager Directions for use Software version 6 XX...
Страница 8: ...4 Symbols Welch Allyn RetinaVue 100 Imager...
Страница 14: ...10 Introduction Welch Allyn RetinaVue 100 Imager...
Страница 59: ...Directions for use Using the RetinaVue 100 Imager 55...
Страница 86: ...82 General compliance and standards Welch Allyn RetinaVue 100 Imager...
Страница 112: ...108 Appendix Welch Allyn RetinaVue 100 Imager...
Страница 114: ......
Страница 115: ......
Страница 116: ...Material No 411492...