Perform an eye exam using the Auto exam mode
Auto exam mode is the default image capture mode.
Available automatic features include:
•
Image capture
•
Focus adjustment
•
Flash brightness adjustment
•
Sequential image capture order of the right and left eye
•
Navigation to the image Inspection screen
○
In addition to automatic image capture, manual capture is also available.
Note
To ensure that the patient's pupils sufficiently dilate to at least 3.5 mm
diameter, adjust the room lighting to the lowest possible level. If
necessary, have the patient sit in a dark room for 5 minutes to dilate their
pupils.
Note
While the RetinaVue 100 Imager can be used on patients with cataracts
and other eye opacities, the use of the RetinaVue 100 Imager may result in
a lower quality image due to the increased reflection of the flash off the
patient's intraocular lens.
•
Ensure that the SD card is installed into the RetinaVue 100 Imager.
•
Ask your patient to remove their glasses, contacts can remain in place.
•
Ensure that the patient sits on the edge of an exam table and stand in front of the
patient to take the image.
Alternatively, ask the patient to sit in a chair and sit directly across from the patient
with your legs together on the same side as the examined eye.
•
Ask the patient to sit up straight and hold their head in a stationary position during
the entire procedure.
48 Using the RetinaVue 100 Imager
Welch Allyn RetinaVue™ 100 Imager
Содержание RetinaVue 100 Imager
Страница 1: ...Welch Allyn RetinaVue 100 Imager Directions for use Software version 6 XX...
Страница 8: ...4 Symbols Welch Allyn RetinaVue 100 Imager...
Страница 14: ...10 Introduction Welch Allyn RetinaVue 100 Imager...
Страница 59: ...Directions for use Using the RetinaVue 100 Imager 55...
Страница 86: ...82 General compliance and standards Welch Allyn RetinaVue 100 Imager...
Страница 112: ...108 Appendix Welch Allyn RetinaVue 100 Imager...
Страница 114: ......
Страница 115: ......
Страница 116: ...Material No 411492...