Removing and refitting the source enclosure
4-9
To remove the source enclosure:
1. Prepare the instrument for working on the source (see
.
2. Remove the probe from the source.
•
If you are removing an ESI probe, see
.
•
If you are removing an IonSABRE II probe, see
.
3. Disconnect the probe adjuster and options cables from the instrument’s
connectors.
4. Pull the source enclosure release (located at the bottom, right-hand side)
outward, and swing open the enclosure.
Warning:
The source components can be contaminated with
biohazardous and/or toxic materials. Always wear
chemical-resistant, powder-free gloves while performing this
procedure.
Warning:
The probe and source can be hot. To avoid burn injuries,
take great care while working with these components.
Warning:
The corona pin tip is sharp. To avoid puncture wounds,
take great care while working with the source enclosure open if a
corona pin is fitted.