3-12
Changing the Mode of Operation
The combined APPI/APCI source components
The combined APPI/APCI source comprises the standard IonSABRE II probe
and a source enclosure with an APPI lamp drive incorporated.
The combined APPI/APCI source enclosure:
The UV lamp, which you ignite via a control in the MassLynx Tune window,
provides a constant photon output. You vary the intensity of incident
radiation upon the sample molecules by adjusting the distance between the
UV lamp and probe tip.
Caution:
To prevent damage to the corona pin and lamp assembly,
ensure that the lamp assembly does not touch the corona pin when the
source enclosure door is closed.
APPI lamp drive assembly