4-92
Maintenance Procedures
Replacing the source assembly seals
To avoid excessive leakage of solvent vapor into the laboratory atmosphere,
the following seals must be renewed at intervals of no greater than 1 year:
•
Probe adjuster assembly probe seal
•
Probe adjuster assembly nebulization gas seal
•
Source enclosure seal
•
Nebulizer gas seal
•
Desolvation gas seal
Removing the probe adjuster assembly probe and source
enclosure seals
Required materials
•
Chemical-resistant, powder-free gloves
•
O-ring removal kit
To remove the probe adjuster assembly probe and source enclosure seals:
1. Remove the source enclosure from the instrument (see
2. Use the O-ring removal kit to carefully remove the following seals from
the probe adjuster assembly (see
•
Probe seal
Warning:
To avoid excessive leakage of toxic and biohazardous
solvent vapor into the laboratory atmosphere, the seals listed
below must be renewed, at intervals of no greater than 1 year,
exactly as described in this section.
Warning:
The source components can be contaminated with
biohazardous and/or toxic materials. Always wear
chemical-resistant, powder-free gloves while performing this
procedure.
Содержание Xevo TQD
Страница 12: ...xii...
Страница 20: ...xx Table of Contents...
Страница 23: ...Uses and compatibility 1 3 Xevo TQD TP03406...
Страница 24: ...1 4 Specifications and Operating Modes Xevo TQD with visor up TP03407...
Страница 42: ...1 22 Specifications and Operating Modes...
Страница 52: ...2 10 Preparing for Operation...
Страница 76: ...3 24 Changing the Mode of Operation...
Страница 194: ...4 118 Maintenance Procedures...
Страница 208: ...A 14 Safety Advisories...
Страница 234: ...B 26 External Connections...
Страница 238: ...C 4 Materials of Construction and Compliant Solvents...
Страница 250: ...Index 8...